`
5. Replace the batteries when the scale displays battery symbol.
6. OVERLOAD WARNING:
7. Remove the weight immediately when the scale displays ’EEEEEE’ or damage
to the scale will occur.
TO MAINTAIN
Clean after use with a lightly dampened cloth (Do not use solvents).
Do not
place on microwave oven or dish washer, or immerse in water.
TO COUNT
1.
To set your scale to parts counting mode, scroll through to weighing by
pressing
CAL/MODE
key when the display shows zero.
2.
Continue to press CAL/MODE key to select the desired parts counting until:
rEF5, rEF1O, rEF2O and rEF5O pieces. After ‘rEF5O’ is displayed, the indicator
will scroll back to gram weighing mode.
3.
With the selected parts counting unit still displayed. Put the desired sample
into the bowl or container. Allow the sample to stabilize for about 5 seconds
then press TARE/PRINT key, “C” appears on the bottom left of the display.
The parts counting.
TO INPUT MEMORY
1. Press TARE/PRINT key to make sure the LCD shows zero.
2. Put the 1 st load on to the platform and get the 1st reading.
3. Press ON/MEM key to store the 1st reading, ‘M’ will appear on the upper left
corner of the LCD.
4. The LCD should show zero now, if not, press TARE/PRINT key.
5. Now put the 2nd load on to the platform and get the 2nd reading.
6. Press ON/MEM key again, ‘M’ disappears and the LCD shows the SUM of the
1st and 2nd readings.
7. Now, you can press ON/MEM key again to store the 1st SUM, LCD should go
back to zero and ‘M’ appears again.
8. Put the 3rd load on to the platform and get the 3rd reading.
9. Press ON/MEM key to get the 2nd SUM (of the 1 st, 2nd and 3rd readings)
10. Go through steps I to 6 above for further Operation.
CALIBRATION INSTRUCTIONS
1. Allow the scale to warm up for about 2 minutes
2. Remove all items from the platform
3. When the LCD shows ‘0.0’ and in the gram mode, press and hold the
CAL/MODE key until the LCD shows ‘500.0’.
4. Now put on to the platform the calibration weight (500g)
5. Wait for about 4 seconds and do not disturb the scale as it is performing
internal calibration.
6. The LCD will show ‘PASS’ if calibration is successful.
7. If LCD shows ‘E’, calibration did not succeed, repeat steps 1 to 6 above.
Tested to comply with FCC standards