DP-10

Air Master

Operation Manual

Summary of Contents for VRN DP-10

Page 1: ...DP 10 Air Master Operation Manual...

Page 2: ......

Page 3: ...ontraindication 11 6 Precautions 11 7 Cleaning and sterilization 13 8 Maintenance 15 9 Storage and transportation 16 10 Accessories list 17 11 Troubleshooting 17 12 After sale service 20 13 Symbols 20...

Page 4: ...se of any discrepancy the binding version is the Chinese text The manual is compiled in accordance with the relevant laws and regulations of China and the specific conditions of the DP 10 manufactured...

Page 5: ...ction photocopying or translation of any content in the manual is not allowed to be translated into other languages The operator must strictly follow this manual to operate Otherwise URIT Medical Elec...

Page 6: ...4 contact Guilin Veirun Medical Technology Co Ltd or an authorized agent...

Page 7: ...laque on the surface of teeth and in periodontal pockets polishing the teeth surface 1 5The product service life is 5 years 2 Technical parameters 2 1 The A weighting sound pressure level produced by...

Page 8: ...and use only dry and clean air 3 2 The recommended pressure of water supply of 0 15 MPa to 0 4 MPa 3 3 The water and air supply at the rear of the Air Master adopts the B type connector according to...

Page 9: ...7 4 Installation 4 1Schematic diagram of the device 1 nozzle 2 handpiece 3 device body 4 polisher container 5 cap 6 connector Figure 1 1 2 3 4 5 6...

Page 10: ...8 4 2 Handpiece installation Connect handpiece by arrow direction See Figure 2 Figure 2...

Page 11: ...9 4 3 Filling the powder Firstly unscrew the cap of container counterclockwise fill the powder in the container and then install the cap clockwise See Figure 3 Figure 3 Pouring the powder...

Page 12: ...10 4 4 Subgingival nozzle installation match the subgingival polisher handpiece Figure 4 4 5 Connection of polisher handpiece and dental scaler...

Page 13: ...cause respiratory difficulties 6 Precautions 6 1 Please read carefully and understand the instructions before operating 6 2 Follow all the warnings and instructions on the device when operating 6 3 Th...

Page 14: ...of air or water leakage and bursting of Air Master 6 8 It must use the dry and clean compressed air and the air compressor water and air filtration systems must be maintained regularly to ensure the...

Page 15: ...sposing and discharging waste 7 Cleaning and sterilization 7 1 When using alcohol for sterilization use 70 80 volume ratio ethanol to soak a piece of clean dry gauze and then use this gauze to wipe th...

Page 16: ...ilization ensure to remove the tip from the handpiece and avoid mixing it with other instruments for sterilization 4 Sterilizable parts can be sterilized at least 250 time 5 It is strictly forbidden t...

Page 17: ...d of the device body and container As the container cap needs to be removed repeatedly once it is damaged or excessively worn the sealing O ring should be replaced in time 8 3 After each use please po...

Page 18: ...d be installed or stored in a cool dry and ventilated place 9 3 Do not mix with toxic corrosive flammable and explosive materials during storage 9 4 The device should be far away from the earthquake s...

Page 19: ...5 mm 2 NOTICE The manual does not exhaustively list the accessories and specifications of the DP 10 Please refer to the random delivery materials and packing list for details 11 Troubleshooting Error...

Page 20: ...er do not escape from the polisher nozzle There are no air or water Check if there is air or water in dental quick connector Handpiece is blocked by powder Pull out the handpiece and check if air or p...

Page 21: ...No powder with air supply There is too little powder in container Check if there is too little powder in the container The supply of air is wet and the powder is caking in container Empty and dry the...

Page 22: ...sponsible for lifetime maintenance Irreparable device damage caused by non designated dedicated maintenance personnel is not covered by the free warranty 13 Symbols Symbol Description Symbol Descripti...

Page 23: ...ption Symbol Description Sterilization at 135 in an autoclave Use by date Atmospheric pressure limitation Humidity limitation Temperature Limit This way up Fragile handle with care Keep away from rain...

Page 24: ...hts The company reserves the right to modify the design technology accessories user manual content and packing list content of the product at any time without notice In case of discrepancies the actua...

Page 25: ...ication sheet of the seals Specification 11x 1 5 Place handpiece connector Specification 6x 1 5 Place handpiece connector Specification rubber seals Place handpiece connector Specification 33x 1 5 Pla...

Page 26: ...he flavor For patients known to suffer such reactions use the powder without flavor provided by VRN 3 The powder containing sodium bicarbonate provided by VRN can be used only for supra gingival appli...

Page 27: ...25...

Page 28: ...lation June 2021 Manufacturer Name URIT Medical Electronic Co Ltd Address No D 07 Information Industry District High Tech Zone Guilin Guangxi 541004 P R China Tel 86 773 2288586 Fax 86 773 2288560 Web...