URCERI

IR-817

Calibration Instruction

Preparations:

1. Ambient temperature: 23

℃±

2

.

2. Place the meter in a constant temperature environment for at least 4 hours.

3. Set the standard temperature source to 300.0

.

4. Calibrating distance fixed at 1 inch.

Calibration steps:

Long press MODE and the trigger button at the same time, and install the battery, the LCD display “----”,

and then loose the buttons to enter into the calibration mode. See below:

Ambient Temperature Calibration Mode

The meter will display the current temperature, press the Flashlight button and the Laser button to

increase or decrease value until the reading is same as the standard value, then press the MODE button to
save the adjusted data. Press the MODE button again to enter into the Ambient Humidity calibration mode.

Ambient Humidity Calibration Mode

After entering into the Ambient Humidity calibration mode, press the Flashlight button and the Laser

button to increase or decrease value until the reading is same as the standard value, then press the MODE
button to save the adjusted data.

Press the MODE button again to enter into the next step.

Ambient Temperature

Calibration Mode

Ambient

Humidity

Calibration Mode

IR

Temperature

Calibration Mode

Type

K

Temp.

Calibration Mode