Page 1: ......
Page 2: ...1 INTRODUCTION 3 Section 3 GENERAL 3 1 UNPACKING INSTALLATION 4 3 2 POWER SUPPLY 4 3 3 CONTROL PANEL 4 Section 4 OPERATION 4 1 INSTRUCTIONS 5 4 2 REMOTE CONTROL CONNECTIONS 6 4 3 RS 232 COMMANDS 7 Section 5 ROUTINE MAINTENANCE 5 1 CLEANING 8 5 2 STORAGE 8 Section 6 SPECIFICATIONS 8 Section 7 ACCESSORIES REPLACEMENT PARTS 9 INDEX 10 1 ...
Page 3: ...ed immediately to Univentor or an authorised representative You must retain the original packing box and contents for inspection by the freight handler Univentor will replace any new instrument damaged in shipping with an identical product as soon as possible after the claim filing date Claims not filed within 30 days after the shipping date will be invalid Do not return damaged goods to Univentor...
Page 4: ... to start stop wait change flow rate and to start or stop external instruments such as the Univentor Microsampler or valves A built in computer makes the unit self calibrating and takes care of all the calculations required to control flow rates and delivered volumes for different sizes and types of syringes For your convenience the setting for the most common syringes are pre programmed and readi...
Page 5: ...ogether with brief instructions are located on the panel of the Univentor 864 Syringe Pump See Figure 4 1 TEXT FUNCTION Fast feed Switch for forward reverse fast feed The carriage moves to the desired position quickly Set Press the right hand arrow to move the cursor for different settings Press the up and down arrow to change the value where the cursor is Syringe Stroke mm Set the length of the s...
Page 6: ...own arrows The injected volume is shown on the display window continuously The pump stops when it has reached the preset volume The units will be the same as the syringe volume For continuous injection leave the preset volume in the SET mode 9 Press Start Stop button to start the pump Present action indicates the mode of the pump The pusher moves forward forcing the plunger into the syringe at the...
Page 7: ...rt on the pump The pump is equipped with TTL logic and a built in RS 232 interface via the 15 pin connector at the rear There is also a relay that can be operated through RS 232 to control equipment outside the pump e g switches for light PIN FUNCTION 1 2 3 Used for RS 232 Computer Control 4 13 START STOP Momentarily applying 5 12 V DC between Pin 4 and Pin 13 will stop the pump if it is running o...
Page 8: ... volume accumulator OK RUN STP to stop Start forward EP ES RUM STP to stop Start forward OK EP ES REV STP to stop Start reverse EP ES REW STP to stop Start reverse OK EP ES FFF STP to stop Fast feed forward EP FRF STP to stop Fast run forward OK EP FFB STP to stop Fast feed backward EP FRB STP to stop Fast run backward OK EP RBC Relay closed OK RBO Relay open OK STP Stop OK SSS Send syringe stroke...
Page 9: ... speed setting Preset speed is controlled via an automatic closed loop speed control Fast feed Pusher movement of 70 mm min both forwards and backwards Syringes 1 8 syringes with stroke length from 40 mm to 90 mm and O D from 6 mm to 32 mm Preset volume 1 1000 of syringe volume Min pusher travel rate 0 006 mm min Max pusher travel rate 6 mm min Max pusher force 200 N Pusher movement accuracy 0 01 ...
Page 10: ...ont 2401028 Syringe holders for 3 syringes up to 10 ml Back 2401025 Syringe holders for 4 syringes up to 2 5 ml Front 2401026 Syringe holders for 4 syringes up to 2 5 ml Back 2401090 Communication cable for RS 232 control of the 802 864 Syringe Pump with 15 to 25 pol 2401091 Communication cable for RS 232 control of the 802 864 Syringe Pump with 15 to 9 pol 2401092 Communication cable for RS 232 c...
Page 11: ... SYRINGES 4 LIMIT SWITCH 5 MAINTENANCE AND STORAGE 8 OPTIONAL ACCESSORIES 9 OPERATING INSTRUCTIONS 5 OUT OF RANGE OOR 6 PACKING LIST 4 POWER SUPPLY 4 REMOTE CONTROL CONNECTIONS 6 RESET 6 RS 232 CABLE 9 RS 232 COMMANDS 7 SERVICE 2 SETTING SYRINGE SIZE 5 SPIN ERR 6 START STOP 5 SYRINGE SIZES 3 SPECIFICATIONS 8 UNPACKING AND INSTALLATION 4 WARRANTY 2 10 ...
Page 12: ...BLB029 Bulebel Industrial Estate Zejtun ZTN 3000 Malta Tel 356 21 895824 Fax 356 21 895835 E mail info univentor com Homepage http www univentor com ...