Section 6 - SPECIFICATIONS

Power 

POWER SUPPLY: 100 - 240V  AC  50 - 60 Hz.

BATTERY: 12V  400 mA.

Dimensions

120 mm (w)  x  285 mm (l)  x 95 mm (h).

Weight

1.8 kg.

Shipping weight

3 kg.

Drive motor

Pulse free DC motor with variable speed setting.

Fast feed

Pusher movement of 45 mm/min.

Syringe

1 glass and gas-tight 10 mL syringe with 60 mm stroke.

Min. liquid flow  rate

0.4 mL/hr.

Max. liquid flow rate

10 mL/hr.

Min. air flow rate

50 mL/min.

Max. air flow rate

999 mL/min.

Min. air pressure/8323001

0.5 bar.

Max. air pressure/8323001

1.0 bar.

Min. air pressure/8323002

0.3 bar.

Max. air pressure/8323002

0.5 bar.

Max. pusher force

100 N.

Pusher movement tolerance

+/- 0.01 mm or +/- 1% of total distance.

Concentration tolerance

+/- 0.15 units of displayed % value.

Display

2x16 Characters.

Safety features

Audible alarm and red LED.

Section 7

ACCESSORIES & REPLACEMENT PARTS

ORDERING INFORMATION

CAT. No.

DESCRIPTION

8323001

Univentor 400 Anaesthesia Unit.

8323002

Univentor 400 Anaesthesia Unit adapted for airpump.

2423017

Gas Outlet Filter.

8433005

Gas Routing Switch

8329001

Induction Chamber 0.8l.

8329002

Induction Chamber 1.4l.

9

Summary of Contents for 400

Page 1: ...THE UNIVENTOR 400 ANAESTHESIA UNIT ...

Page 2: ...ion 3 GENERAL 3 1 UNPACKING INSTALLATION 4 3 2 POWER SUPPLY 4 3 3 CONTROL PANEL 4 3 4 DISPLAY 5 3 5 SAFETY FEATURES 5 Section 4 OPERATION 4 1 INSTRUCTIONS 5 Section 5 ROUTINE MAINTENANCE 5 1 CLEANING 7 5 2 LEAKAGES 7 5 3 REPLACING PARTS 7 5 4 STORAGE 7 5 5 TROUBLE SHOOTING 7 Section 6 SPECIFICATIONS 9 Section 7 ACCESSORIES REPLACEMENT PARTS 9 INDEX 10 1 ...

Page 3: ...rted immediately to Univentor or an authorised representative You must retain the original packing box and contents for inspection by the freight handler Univentor will replace any new instrument damaged in shipping with an identical product as soon as possible after the claim filing date Claims not filed within 30 days after the shipping date will be invalid Do not return damaged goods to Univent...

Page 4: ...s the Univentor 8323002 has been adapted to work with a pulse free airpump Air and anaesthetic is mixed in a vaporiser according to settings and may be delivered into an anaesthetised box through a mask or to a ventilator Your unit is calibrated for Isoflurane and must at all times be operated with a 10 mL stroke 60 mm gas tight glass syringe Should you wish to use other liquid anaesthetics you wi...

Page 5: ...structions are located on the panel of the Univentor 400 Anaesthesia Unit See Figure 3 3 TEXT FUNCTION Fast Feed forward reverse fast feed to position pusher Set increasing decreasing liquid flow rate Start Stop starting stopping liquid delivery Lights green when delivering and red in case of alarm ref 3 5 4 FIGURE 3 3 The control panel of the Univentor 400 Anaesthesia Unit Fast Feed Set Start Sto...

Page 6: ...ely 150 250 mL min Rats approximately 400 500 mL min Should you use a chamber you will need to increase the air flow and anaesthetic until the animal is asleep Mice approximately 400 500 mL min Rats approximately 700 800 mL min Once asleep turn the stopcock and route the gas to the mask and decrease the airflow NOTE the unit will NOT operate without air 5 Section 4 OPERATION 3 4 DISPLAY 1 AIR FLOW...

Page 7: ...inter is in the required stopping position 9 Remove protective cover and connect one end of the supplied PVC tubing to the Gas Outlet and the other to the anaesthetised box mask or ventilator Keep tubing as short as possible 10 Calculate the operating time available by dividing the liquid volume with the flow rate and make sure that the time is adequate Flow rate is displayed when the instrument i...

Page 8: ...n display contact your Univentor representative 5 3 REPLACING PARTS We recommend that the gas outlet filter part number 2423017 is changed at least yearly or as necessary according to use 5 4 STORAGE If the Univentor 400 Anaesthesia Unit is not to be used for a significant length of time it is recommended to clean the instrument fit protective covers on air inlet and outlet and store it safely in ...

Page 9: ...blocking the liquid 4 Animal not being anaesthetised a Faulty air sensor contact your Univentor representative b Syringe no longer gas tight replace c Cannula to vaporiser is blocked to test fill syringe with air connect to luer lock and try pushing plunger If air passes easily it is not blocked If very tough to push and plunger bounces back when released it is blocked To unblock remove gas outlet...

Page 10: ...mL min Min air pressure 8323001 0 5 bar Max air pressure 8323001 1 0 bar Min air pressure 8323002 0 3 bar Max air pressure 8323002 0 5 bar Max pusher force 100 N Pusher movement tolerance 0 01 mm or 1 of total distance Concentration tolerance 0 15 units of displayed value Display 2x16 Characters Safety features Audible alarm and red LED Section 7 ACCESSORIES REPLACEMENT PARTS ORDERING INFORMATION ...

Page 11: ...HETIC 6 CONTROL PANEL 4 DAMAGED SHIPMENTS 2 DISPLAY 5 LEAKAGES 6 LIMIT SWITCH 6 LIQUID FLOW RATE 5 6 MAINTENANCE AND STORAGE 7 OPERATING TIME 6 OPERATING INSTRUCTIONS 5 6 PACKING LIST 4 POWER SUPPLY 4 REPLACING PARTS 7 SERVICE 2 SPECIFICATIONS 9 SYRINGE 3 5 TOLERANCES 9 TROUBLE SHOOTING 7 8 UNPACKING AND INSTALLATION 4 WARRANTY 2 10 ...

Page 12: ...BLB029 Bulebel Industrial Estate Zejtun ZTN 3000 Malta Tel 356 21 895824 Fax 356 21 895835 E mail info univentor com Homepage http www univentor com ...

Reviews: