background image

14

IMPORTANT

It is important 

to perform 

calibration and 

subsequent 

measurements in 

solution with the 

same temperature 

and salinity. 

FIELD 

CALIBRATION

For field 

calibrations, please 

see alternative 

calibration 

procedure below.

IMPORTANT

Vigorous bubbling 

water with any 

gas may cause 

the water to cool 

considerably. 

Monitor the 

temperature to find 

a suitable bubbling 

rate, which does 

not cool the water 

significantly. 

Please read the entire procedure below before the calibration is 
commenced . It is best to first prepare the Na

2

S stock solution so 

it is ready . Then prepare the buffer and perform the calibration 
quickly thereafter in order to avoid reintroduction of oxygen .

NOTE: For field calibrations, an alternative method is 
recommended - this will enable you to prepare reagents to bring 
into the field for easy calibration . 

Laboratory calibration

1 .  Prepare a stock solution
A stock solution of S

2-

 (≈ 0,01M total sulphide) is prepared 

anaerobically by dissolving 0,24 g Na

2

S * 9 H

2

O in 100 mL of 

N

2

-flushed water in a closed container . The final concentration of 

the stock solution should be determined by standard analysis (e .g . 
Cline 1969 or Budd & Bewick 1952) . Keep the stock solution out of 
contact with oxygen .

2 .  Prepare the calibration buffer
Any standard pH buffer, with a pH value less than 4, can be used . 
Remove oxygen from a volume of the buffer . This can be done in 
two ways:

a .  By vigorously bubbling with an 

oxygen-free inert gas (e .g . N

2

) for at 

least 5 minutes . NOTE: Vigorously 
bubbling buffer with any gas may cause 
the water to cool considerably . Monitor 
the temperature to find a suitable 
bubbling rate, which does not cool the 
buffer significantly . 

b .  By adding a suitable reductant (e .g . 

Ti(III)Cl; MERCK supplies this in a 10% 
HCl solution) to the oxygen free buffer 
to a final concentration of 50 mM . Add 
a few glass beads (2-3 mm in diameter) 
to facilitate mixing . The transfer is 

Summary of Contents for SULF Series

Page 1: ...1 HYDROGEN SULFIDE SENSOR USER MANUAL...

Page 2: ...Hydrogen sulfide sensor manual Copyright 2018 Unisense A S Version March 2018...

Page 3: ...HYDROGEN SULFIDE SENSOR MANUAL UNISENSE A S...

Page 4: ...ing a new sensor 11 Polarization 11 Pre polarization 12 Calibration 13 Calibration theory 13 Calibration in the laboratory 14 Calibration in the field 16 Re calibration 18 MEASUREMENTS 19 Mounting the...

Page 5: ...d repair or modification of the product In no event will Unisense A S be liable for any direct indirect consequential or incidental damages including lost profits or for any claim by any third party a...

Page 6: ...orking day E mail sales unisense com Unisense A S Tueager 1 DK 8200 Aarhus N Denmark Tel 45 8944 9500 Fax 45 8944 9549 Further documentation and support is available at our website www unisense com CO...

Page 7: ...7 RECOMMENDED AMPLIFIERS One channel amplifier Microsensor Monometer Multi channel amplifiers Microsensor Multimeter Field Microsensor Multimeter...

Page 8: ...e and minimal stirring sensitivity the H2S sensors makes it possible to make reliable and fast measurements at high spatial resolution Our H2S microsensors are miniaturized amperometric sensors with i...

Page 9: ...sensor signal is generated by re oxidation of ferrocyanide at the anode in the tip of the sensor Jeroschewski et al 1996 The internal guard anode facilitates a constant ratio of ferri to ferrocyanide...

Page 10: ...IMPORTANT Wrong polarization may destroy the sensor Type I hydrogen sulfide sensors label on the sensor shaft reads SULF xxx xxxxxx should be polarised with 200 mV Type II hydrogen sulfide sensors lab...

Page 11: ...it must be pre polarized for at least 2 hours before it can be calibrated and used After shorter periods without polarization the sensor should be pre polarized until it has shown a stable signal for...

Page 12: ...2S Stot 2 1 H30 for pH 9 Jeroschewski et al 1996 Thus it is necessary to know the pH i e to know H3O of the sample calibration solution to calculate Stot 2 see Figure 1 pH log H In solutions with a pH...

Page 13: ...l has stabilized during pre polarization The H2S microsensor responds linearly over a certain range e g 0 300 M above which the slope of the response curve decreases but the response and resolution ab...

Page 14: ...alibration 1 Prepare a stock solution A stock solution of S2 0 01M total sulphide is prepared anaerobically by dissolving 0 24 g Na2S 9 H2O in 100 mL of N2 flushed water in a closed container The fina...

Page 15: ...mes Mix the solutions If you have not added reductant oxygen that dissolve in the calibration solution will oxidize the sulfide so the stock solution should be added immediately after the oxygen remov...

Page 16: ...reagent 2 into a 10 ml glass vial with gas tight membrane lid Avoid head space Prepare as many vials as the number of calibrations you expect to make Bring reagent 1 the vials with reagent 2 and 1 ml...

Page 17: ...may require calibration only every 24 hours or less To minimize the need for calibration keep the sensor polarized between measurements if possible unless the time between measurements exceeds several...

Page 18: ...trical fields may interfere with the sensor signal Therefore we recommend that unnecessary electrical mechanical equipment is switched off and that the sensor or wires are not touched during measureme...

Page 19: ...red microns The sensor has been painted black to reduce this phenomenon however light may still enter the sensor through the tip which is not possible to paint Micromanipulator Interferences for gases...

Page 20: ...ts The most frequent construction options are described under sensor specification at our website The options include e g customer specified dimensions cable length etc If your specifications for a sp...

Page 21: ...he tip exposed to water or air The room in which the H2S microsensor is stored should be dry and not too hot about 5 35 C Type II Store dry on shelf at room temperature Cleaning the sensor Standard pr...

Page 22: ...1952 Photometric determination of hydrogen Sulfide and reducible sulfur in alkalies Anal Chem 24 1536 1540 Cline J D 1969 Limnol and Oceanogr 14 454 458 Jeroschewski P Steuckart C K hl M 1996 Anal Ch...

Page 23: ...the sensor tip Solution Rinse with 96 ethanol rinse with 0 01 M HCl and rinse with water Problem Measurements in two environments with equal H2S concen trations exhibit different signals Possible caus...

Page 24: ...UNISENSE DENMARK www unisense com info unisense com...

Reviews: