5
■
Graph trigger start level
If you select "Continued trigger" or "Single trigger" in the graph function setting, set the graph
trigger start level.
≪
Setting value
≫
-99999 to 99999
How to set
■
Graph trigger slope
If you select "Continued trigger" or "Single trigger" in the graph function setting, set the graph
trigger slope conditions.
≪
Setting value
≫
Either, Rising, Falling, Beyond, Below
How to set
- Either:
Starts plotting a graph when the torque crosses the trigger level.
- Rising:
Starts plotting a graph when the torque crosses the trigger level from a lower
value to an upper value.
- Falling:
Starts plotting a graph when the torque crosses the trigger level from an upper
value to a lower value.
- Beyond:
Starts plotting a graph when the torque is greater than the trigger level.
- Below:
Starts plotting a graph when the torque is less than the trigger level.
5-15.
Key LOCK
Disables all key operation on the measurement screen
(comparison, graph) while the external input B5
terminal is short-circuited.