22
4
CALIBRATION
22
C
A
LIB
R
A
TIO
N
Chapter
4
4.
Select the increment and determine with
the [OK] button.
4-8.
Digital Zero Limit
(optional if there is no change)
This is a load limit to allow digital zero.
When digital zero is executed, if the difference from the zero calibration point is larger than the
setting value, an error will result, and the indicated value will not be zeroed.
≪
Setting value
≫
0 to 99999
How to set
1.
Press the [SET.] button. (see P.17)
2.
Press the [Calibration] button. (see P.19)
3.
Select page 2, and press the [DZ Limit]
button.
4.
Enter the digital zero limit value by the
numerical keys and determine with the
[OK] button.
SET.
Calibration
Page 2
→
→