Hold Function
27
Hold Function
Chapter
6
Start level (START LEV.)
Sets the level to start detection.
<Setting range>
-9999 to 9999
Stop level (STOP LEVEL)
Sets the level to stop detection.
<Setting range>
-9999 to 9999
Zero clear at the start of hold section (START CLR)
Set whether or not to reset the angle to zero at the start of hold section.
* This function works only when the peak hold (angle) is selected.
<Setting range>
OFF, ON
- OFF:
The angle is not reset to zero at the start of hold section.
- ON:
The angle is reset to zero at the start of hold section.
* Do not use this function in combination with "zero clear from the recording start (REC.ST.CLR)"
that is described in encoder setting.
6-2.
Sample hold
Holds an arbitrary point.
- Operation of sample hold
Holds the indicated value at
key down.
Press the
key again to release hold.
key input
Time
Torque
Indicated value
HOLD display
Hold section
HOLD
Lights off
Lights on
Lights off
HOLD
HOLD