6.FUNCTION SETTING
39
6-1-5. Zero tracking
This function automatically adjusts slow drifts and slight shifting of the zero point due to small
amounts of accumulation on a scale.
Operation
ZT PEROPD:
ZT RANGE:
- Zero tracking is the function to reset gross weight to zero automatically when the
state remaining the zero point drift within the zero tracking range continues more
than set period.
- Zero tracking period must be set between 0.1 ~ 9.9 second and its range must be
set between 1 ~ 99 using 1/4 resolution interval of weight display.
For instance, 02 corresponds to 0.5 and 12 corresponds to 3 of weight display.
Zero tracking range is not proportional to the Minimum scale division regardless of
the setting state of Minimum scale division.
Zero tracking does not work if you set the period to 0.0 sec. and range to 00.
Boundary of zero track
Tracking period
Weight value
Tracking range
+
0
-
Tracking period
From the point when it returned within the range,
counting will be resumed.
MODE
OPERATION
→
→
Press
ZERO TRACKING (PERIOD)
→
Input 0.1 ~ 9.9
OK
→
MODE
OPERATION
PAGE
→
→
Press
ZERO TRACKING (RANGE)
→
→
Input 1 ~ 99
OK
→