5.CALIBRATION
32
5-4-6. Minimum scale division
The minimum scale division can be from 001 to 100.
The display resolution obtains from dividing minimum scale division by capacity.
Operation
The display resolution ignores the decimal place.
For example : When minimum scale division is 001 then
if capacity is 10.000, the display resolution is 1/10000;
if capacity is 100.00, the display resolution is 1/10000;
if capacity is 40.000, the display resolution is 1/40000.
5-4-7. Balance weight
The load value (Balance weight) that is applied to the scale for calibration. The input range of Balance
is between 0 to 99999.
Operation
5-4-8. Gravitational acceleration
Slight error may occur if the scale moved from the location of calibration due to gravitational changes.
If the scale is used in the location of calibration, this function is not necessary. But after calibration the
scale will ship to different location, before Span calibration do operation procedure as follows.
- Select GRAV. ACCELERATION , input relative number to located Gravitational Acceleration.
- Do Span calibration.
- Input the number following new place Gravitational Acceleration.
Operation
Gravitational accceleration
MODE
CALIBRATION
MIN.SCALE DIVISION
→
→
Press
→
Input 1 ~ 100
OK
→
MODE
CALIBRATION
→
→
Press
BALANCE WEIGHT VALUE
→
Input 0 ~ 99999
OK
→
01
9.806
02
9.805
03
9.804
04
9.803
05
9.802
06
9.801
07
9.800
08
9.799
09
9.798
10
9.797
11
9.796
12
9.795
13
9.794
14
9.793
15
9.792
16
9.791
MODE
CALIBRATION
PAGE
→
→
Press
GRAV.ACCELERATION
→
→
Input 01 ~ 16
OK
→