CALIBRATION
25
C
A
LI
B
R
A
TI
O
N
Chapter
4
4-3.
Preparation before calibration
■
LOCK release
Calibration values and setting values can be locked so as not to be changed by misoperation.
In LOCK, there is software LOCK applied by setting. To perform calibration, release the lock.
1.
Select setting mode 3-6(LOCK).
2.
Set the setting LOCK2 to OFF.
Now, the LOCK is released. Upon completion of calibration, apply the LOCK to protect the
calibration values.
■
Unit display (setting mode 4-5)
Set the unit for weighing.
Select the unit from 0: None/1: t/2: g/3: kg/4: lb.
■
Decimal place (setting mode 4-5)
Set the decimal place common to weight-related displays and setting items, etc., by this selection.
Select the decimal place from 0: None/1: 0.0/2: 0.00/3: 0.000.
LOCK2
0: OFF
1: ON
LOCK
For the LOCK and setting values protected, see "11-1.List of setting values" on
page 97.
Point
Unit
0: None
3: kg
1: t
4: lb
2: g
Display selection
Display selection
Decimal place
0: None
1: 0.0
2: 0.00
3: 0.000
In the F701-P, the decimal place is fixed for all other than weight-related data.
* It cannot be changed.
Point