Chapter
11
■
Calibration error display
(Calibration error)
The initial tare exceeds the zero calibration range of the F701-P. Check the load cell for
unnecessary load.
If "
" is displayed under a normal load, there is a need to connect resistance between
the +EXC and -SIG terminals of the load cell and redo zero calibration after shifting the zero point.
The relationships between connected resistance values and input signals are shown below.
-
The numerical values in this table assume that one 350Ω load cell is used.
- The temperature coefficient of the resistance connected here directly affects the accuracy of the
indicator.
Use resistance of 50ppm/ C or higher (approx. 5ppm/ C is recommended).
Resistance value
Input conversion strain
Approximate value
Calculated value
μ-STRAIN
mV/V
875
kΩ
866
kΩ
200
0.1
437
kΩ
442
kΩ
400
0.2
291
kΩ
294
kΩ
600
0.3
219
kΩ
221
kΩ
800
0.4
175
kΩ
174
kΩ
1000
0.5
146
kΩ
147
kΩ
1200
0.6
125
kΩ
124
kΩ
1400
0.7
109
kΩ
110
kΩ
1600
0.8
97
kΩ
97.6
kΩ
1800
0.9
87.3
kΩ
86.6
kΩ
2000
1.0
79.4
kΩ
78.7
kΩ
2200
1.1
72.7
kΩ
73.2
kΩ
2400
1.2
67.1
kΩ
66.5
kΩ
2600
1.3
62.3
kΩ
61.9
kΩ
2800
1.4
58.2
kΩ
57.6
kΩ
3000
1.5
54.5
kΩ
54.9
kΩ
3200
1.6
51.3
kΩ
51.1
kΩ
3400
1.7
48.4
kΩ
48.7
kΩ
3600
1.8
45.9
kΩ
46.4
kΩ
3800
1.9
43.6
kΩ
43.2
kΩ
4000
2.0
41.5
kΩ
41.2
kΩ
4200
2.1
39.6
kΩ
39.2
kΩ
4400
2.2
37.9
kΩ
38.3
kΩ
4600
2.3
36.3
kΩ
36.5
kΩ
4800
2.4
34.8
kΩ
34.8
kΩ
5000
2.5
Connect resistance
bEXC and -SIG.
+EXC
+SIG
-EXC
-SIG