15.BCD DATA OUTPUT (OPTION)
53
15-4. Signal Output Timing
- P.C
Note: P.C should also be acquired in synchronization with the strobe signal as in the case of
the data signal. (Refer to page 53.)
●
HI/LO limit output
ON at stable-time, OFF at unstable-time.
●
In the discharge control mode; simple comparison mode
The signal is output at each ON edge of the completion signal. The value of the number of judging
times is ignored.
●
In the discharge control mode; sequence mode
The signal is output at the ON edge of the completion signal at the intervals of judging times.
- OVER
The signal is output with +ADC OVER (+LOAD) or -ADC OVER (-LOAD) overflow (NET OVER or
GROSS OVER).
- STROBE
Strobe pulses are output in synchronization with BCD data. For reading the data, use the pulse rising
edge(ON
→
OFF). The BCD data update rate can be changed by setting.
Note: The logic on the software varies depending on the equipment on the receiving side.
(0
→
1, 0
→
1)
BCD data
P.C
ON
BCD data
P.C
ON
BCD data
ON
OFF
STROBE
(DUTY about 50%)
If the update rate is changed, the strobe pulse width is also changed proportionally.