- HOLD
At hold-off time, the held value is automatically printed to the UNIPULSE printer coupled with the
F377A through the SI/F. At hold-off time, hold values and real time values are output in the
GROSS area and NET area, respectively. Others are the same as in the setting of "OFF".
Setting range: OFF, MD, HOLD
How to set
5-9.
Indicate Color
The indicated value display color can be changed. By setting Comp.(Comparison Result), the
indicated value display color changes following the comparison status.
OK
green
HI, LO
yellow
HH, LL
red
Setting range: Yellow, Green, Blue, Comp.
How to set
5-10.
Comparison Functions
By the comparison function, the HI limit and LO limit values are set, and when the indicated value
exceeds the HI limit, the HI output is turned on, and when the indicated value falls below the LO
limit, the LO output is turned on. Also, HH limit and LL limit values may be set outside the HI-LO
limit comparison. When the indicated value exceeds the HH limit, the HH output is turned on, and
when the indicated value falls below the LL limit, the LL output is turned on. When the HI, HH, LO
and LL outputs are all off, the OK output is turned on.
〈
HI/LO output conditions
〉
HI:
Indicated value
>
HI limit value
LO:
Indicated value
<
LO limit value
〈
HH/LL output conditions
〉
HH:
Indicated value
>
HH limit value
LL:
Indicated value
<
LL limit value
〈
OK output conditions
〉
OK:
All conditions of HH, HI, LO and LL are off.
SET.
Operation
Page 1
→
→
System Setting
→
SET.
Operation
Page 2
→
→
System Setting
→
Time
Indicated value
HI Limit
LO Limit
HI
LO
OFF
ON
OFF
ON
HH Limit
LL Limit
OFF
ON
OFF
ON
HH
LL
OFF
ON
OK
Summary of Contents for F377A
Page 1: ...18JUL2012REV 3 06 DIGITAL INDICATOR F377A OPERATION MANUAL ...
Page 5: ...What is RoHS 4 M E M O ...
Page 12: ...11 11 STANDARD ...
Page 13: ...12 12 M E M O ...
Page 94: ...93 93 EXPANSION ...
Page 95: ...94 94 M E M O ...
Page 97: ...Expansion Function 96 M E M O ...
Page 152: ......