2
INSTALLATION & CONNECTION
24
INSTALLATION & CONN
ECTION
Chapter
2
■
Connecting to cage clamp terminal block
The output terminal D/A option is using the cage clamp system terminal stand. Please connect in
the following procedure.
1.
Strip the casing 5 - 6mm on the cable to be
connected.
2.
Twist the bare wire to fit the terminal hole.
3.
Insert the supplied screwdriver into the
upper hole and lift upward.
4.
Insert the twisted wires into the lower hole.
5.
Make sure cable is clamped securely and
does not come out with a slight tug.
5 to 6mm
CAUTION
- Cable can be from 24 to 14AWG (0.2 to 2.5mm
2
)
- It is not necessary to solder the cable wires or to fix a solderless terminal.
- If several cables to be inserted to the same hole, twist those cable wires
together and insert.
- If you connect a cable (sensor(s), SI/F, external input and output),
please turn off and be sure to perform the power supply of a main part.