8
SUPPLEMENTS
147
SUPP
LEMENTS
Chapter
8
■
System setting
Protect
○
:Work setting protect,
◎
:System setting protect,
●
:Calibration protect,
レ
:Expansion protect
Memory
S:SRAM, N:NOVRAM
Operation setting Standard
Operation setting Expansion
-Digital filter character
CR HPF / LPF-
*)
At 100Hz or less, setting can be made in units of hertz.
At 100Hz or more, setting can be made in units of 10 hertz.
RS-232C setting Standard
Page No.
Item
Initial value
Setting range
Memory Protect
Command
No.
COM.
Read/
Write
1
1
Digital Filter
OFF
0: OFF
,
2 to 999 Times
N
◎
501
R/W
2
Analog Filter
2: 300Hz
0: 30
1: 100
2: 300
3: 1000 [Hz]
N
◎
502
R/W
3
Backlight
ON Time:
10min
0 to 99 Min (Always ON when "0".)
N
◎
503
R/W
Bright
→
Dark
:
0 min
0 to 99 Min (Always Bright when "0")
*
517
4
Language
1:ENG(
英)
0: JPN(
日
)
1:ENG(
英)
N
◎
504
R/W
5
SI/F Print Out
OFF
0: OFF
1: MD
2: HOLD
N
◎
505
R/W
2
6
Motion Detect
0.0 sec /
0count
0.0 to 9.9 sec / 0 to 99 count
N
◎
506/
507
R/W
7
Zero Tracking
0.0 sec /
0count
0.0 to 9.9 sec / 0 to 99 count
N
◎
508/
509
R/W
8
Vol. Out Filter
1: ON
0: OFF
1: ON
N
◎
510
R/W
9
Indicate Color
Yellow
0: Yellow 1: Green 2: Blue 3: Comparison
N
◎
511
R/W
10 B5 Function Selection
0: WORK 3
0: WORK 3
1: DSP_LOCK
2: KEY_LOCK
N
◎
512
R/W
3
11 B6 OFF Det. Wait
0.00 sec
0.00 to 1.00 sec
N
◎
513
R/W
12 B8 OFF Det. Wait
0.00 sec
0.00 to 1.00 sec
N
◎
514
R/W
13 Measurement Work Selection
1: EXT.
0: COM.
1: EXT.
N
◎
515
R/W
14 Control Input Selection
1: EXT.
0: COM.
1: EXT.
N
◎
516
R/W
15 Password
1
1
Digital Filter
1Hz
1 to 990Hz
*
N
◎
501
R/W
2
Analog Filter
2: 300Hz
0: 30
1: 100
2: 300
3: 1000 [Hz]
N
◎
502
R/W
3
Backlight
ON Time:
10min
0 to 99 Min (Always ON time when "0".)
N
◎
503
R/W
Bright
→
Dark
:
0 min
0 to 99 Min (Always Bright when "0")
*
517
4
Language
1:ENG(
英)
0: JPN(
日
)
1:ENG(
英)
N
◎
504
R/W
5
SI/F Print Out
OFF
0: OFF
1: MD
2: HOLD
N
◎
505
R/W
2
6
Motion Detect
0.0 sec /
0count
0.0 to 9.9 sec / 0 to 99 count
N
◎
506/
507
R/W
7
Zero Tracking
0.0 sec /
0count
0.0 to 9.9 sec / 0 to 99 count
N
◎
508/
509
R/W
8
Vol. Out Filter
1: ON
0: OFF
1: ON
N
◎
510
R/W
9
Indicate Color
Yellow
0: Yellow 1: Green 2: Blue 3: Comparison
N
◎
511
R/W
10 B5 Function Selection
0: WORK 3
0: WORK 3
1: DSP_LOCK
2: KEY_LOCK
N
◎
512
R/W
3
11 B6 OFF Det. Wait
0.00 sec
0.00 to 1.00 sec
N
◎
513
R/W
12 B8 OFF Det. Wait
0.00 sec
0.00 to 1.00 sec
N
◎
514
R/W
13 Measurement Work Selection
1: EXT.
0: COM.
1: EXT.
N
◎
515
R/W
14 Control Input Selection
1: EXT.
0: COM.
1: EXT.
N
◎
516
R/W
15 Password
1
1
Communication Mode
0: Normal
0: Normal
1: Continue
2: Print
N
◎
601
R
2
Baudrate
1: 19200bps
0: 9600
1: 19200
2: 38400
3: 57600
[bps]
N
◎
602
R
3
Data Bit
1: 8bit
0: 7bit
1: 8bit
N
◎
603
R
4
Stop Bit
0: 1bit
0: 1bit
1: 2bit
N
◎
604
R
5
Parity Bit
1: EVEN
0: NONE
1: EVEN
2: ODD
N
◎
605
R
2
6
Delimiter
0: CR
0: CR
1: CR+LF
N
◎
606
R
7
Flow Control
0: OFF
0: OFF
1: RTS/CTS
N
◎
607
R
8
9
10
Summary of Contents for F377A
Page 1: ...18JUL2012REV 3 06 DIGITAL INDICATOR F377A OPERATION MANUAL ...
Page 5: ...What is RoHS 4 M E M O ...
Page 12: ...11 11 STANDARD ...
Page 13: ...12 12 M E M O ...
Page 94: ...93 93 EXPANSION ...
Page 95: ...94 94 M E M O ...
Page 97: ...Expansion Function 96 M E M O ...
Page 152: ......