2
EXP. COMPARISON FUNCTIONS
118
EXP
. COMPARISON
FUNCTIONS
Chapter
2
■
Before value regulation
* For judging that the difference between the previous and present hold values is too large for
comparison, set the Before Value Regulation. If the absolute value of the difference is larger than
this regulation, the Before Value is not renewed. (Not the present hold value but the previous hold
value is used for the next comparison.)
- Before Value Regulation
0 - 99999
Initial value: 99999
2-2.
Before Value Comparison in Double Hold
Before-value-comparison-specific measurement screens in double hold appear.
Also, the Comparison Setting menu is dedicated.
■
Comparison standard
In double hold, Before Value Comparison can be made only on either A or B by selecting either
Hold A or Hold B as Comparison Standard.
- Comparison Standard Hold A / Hold B
Initial value
Hold A
Difference value when the Comparison Standard is Hold A(B) =
Present hold value A(B) - Previous hold value A(B)
FCN-365P024-AU/FCN-360C024-B
External I/O
A1
-
COM1
COM2
-
B1
A2
OUT
DF-HI
WORK0
IN
B2
A3
OUT
HI
WORK1
IN
B3
A4
OUT
OK
WORK2
IN
B4
A5
OUT
LO
WORK3(LOCK)
IN
B5
A6
OUT
DF-LO
T/H
IN
B6
A7
-
COM1
COM2
-
B7
A8
OUT
HOLD END(H/E)
SECTION
IN
B8
A9
OUT
RUN
D/Z
IN
B9
A10
OUT
EVENT
GRAPH TRIG
IN
B10
A11
OUT
SIF
CAL0
IN
B11
A12
OUT
SIF
CAL1
IN
B12
Dif.
Hold B
Hold A
Hold B
Hold A
Hold B
Summary of Contents for F377A
Page 1: ...18JUL2012REV 3 06 DIGITAL INDICATOR F377A OPERATION MANUAL ...
Page 5: ...What is RoHS 4 M E M O ...
Page 12: ...11 11 STANDARD ...
Page 13: ...12 12 M E M O ...
Page 94: ...93 93 EXPANSION ...
Page 95: ...94 94 M E M O ...
Page 97: ...Expansion Function 96 M E M O ...
Page 152: ......