If Relative Minimum or Relative Maximum is selected to Hold, the
following setting items are displayed.
- Detection Start Load
Set the load to start the detection of Relative Maximum (Relative
Minimum).
If you do not use Detection Start Load, press
.
Operation
Setting range (-9999 to 9999)
Main screen
→
Setting
→
Work Setting
→
Hold Setting
→
Use Hold
→
Rel. Max. (or Rel. Min.)
→
Start Load
- Detection Load Difference
Set the load difference of Relative Maximum or Relative Minimum.
Select with the Up key for the Relative Maximum, and with the
Down key for the Relative Minimum.
Operation
Setting range (1 to 19998)
Main screen
→
Setting
→
Work Setting
→
Hold Setting
→
Use Hold
→
Rel. Max. (or Rel. Min.)
→
Load Difference
- Detection Rate
Set the rate to detect Relative Maximum (Relative Minimum).
Relative Maximum (Relative Minimum) is detected when the
detection rate is exceeded based on the relative minimum.
Setting range
Operation
(1/4, 1/2, 3/4, 1, 1.25, 1.5, 1.75, 2, 3, 4 Times)
Main screen
→
Setting
→
Work Setting
→
Hold Setting
→
Use Hold
→
Rel. Max. (or Rel. Min.)
→
Rate
- Detection Ordinal
Set the number of points to be counted before judgment of Relative
Maximum (Relative Minimum).
Operation
Setting range (1 to 15 Times)
Main screen
→
Setting
→
Work Setting
→
Hold Setting
→
Use Hold
→
Rel. Max. (or Rel. Min.)
→
Ordinal
Summary of Contents for DeviceNet F381A
Page 1: ...15APR2013REV 3 10 DYNAMIC FORCE PROCESSOR F381A OPERATION MANUAL ...
Page 9: ...Contents VIII Contents VIII M E M O ...
Page 34: ...25 2 INSTALLATION AND CONNECTION 25 INSTALLATION AND CONNECTION Chapter 2 M E M O ...
Page 147: ...138 8 SPECIFICATIONS 138 SPECIFICATIONS Chapter 8 8 2 Outside dimensions Unit mm ...
Page 164: ......