ATMOS Operation Manual
page 22/24
ATMOS Technical Specifications
Maximum accuracy error formula:
±
𝐸𝐸
= (4%
∗ 𝑑𝑑𝑑𝑑
) + 0.1 [µ
𝑙𝑙
]
P = nominal pipette volume
Absolute accuracy
(= trueness)
Repeatability
(= precision)
P=1000
1000 µl
±4 µl
±1.5 µl
P=200
200 µl
±0.1 µl
±0.3 µl
P=100
100 µl
±0.5 µl
±0.15µl
P=20
20 µl
±0.18 µl
±0.05µl
P=10
10µl
±0.14 µl
±0.06µl
Figure: Atmos accuracy in function of tested volume displacement
Technical Specifications
Dimensions
12.5 x 8.5 x 3 cm
Weight
420 g
Power supply
5V DC, 2A
Battery
2000 mAh, Lithium-Polymer
Operational temp.
from 10 °C to 40°C
Altitude
from sea level up to 3500m