UNION Instruments GmbH
INCA Analyzer Series
– Instruction Manual
Page 28 of 59
System main menu
4.5
Parameters
4.5.1
Gas cooler temperature
Set the temperature of the gas cooler (only available if gas cooler is installed).
4.5.2
EC measurement cycle
The measuring cycle of electrochemical sensors is set using this dialog. “1” means that on
every measuring phase the cycle runs. “2” means that only every other measuring phase the
cycle for EC sensor runs and the gas is measured.
Parameters
Gas cooler temp
EC meas. cycle
Under
Parameters
menu, press
▲ and ▼ to highlight
EC meas. cycle
and press
MENU
/ .
Use
▲ and ▼ to select the cycle‟s setting.
Press
MENU
/
to store setting
EC meas. cycle
1
Saving...
OK
Parameters
Gas cooler temp
EC meas. cycle
Under
Parameters
menu, press
▲ and ▼ to highlight
Gas cooler temp
and press
MENU
/ .
Use
▲ and ▼ to give temperature value of the gas
cooler.
Press
MENU
/
to store setting
Gas cooler temp
1
2.0 C
Saving...
OK