UNION Instruments GmbH
INCA Analyzer Series
– Instruction Manual
System main menu
Page 27 of 59
4.4.3
Calibration purge gas (on/off)
The „calibration after purge gas‟-option is activated by setting this menu
on
or
off
.
4.4.4
Output data (on/off)
The menu activates the
unidirectional
interval data output using the RS-232 interface. The
output data is always transmitted with a baud rate of 9600 bits/s, regardless of the
„Communication‟ setting. Nevertheless it is mandatory to turn this setting „off‟ when using the
H-Bus protocol to communicate with the analyzer and to read its measurement data.
4.4.5
Communication
The menu provides the communication option of the RS-232 connectivity for
bidirectional
communication:
1. INCACtrl
– For INCA configuration tool and H-Bus protocol working at 115200 bits/s
2. H-Bus (9600)
– H-Bus protocol working at 9600 bits/s
3. H-Bus (2400).
– H-Bus protocol working at 2400 bits/s
Further information can be found under the Chapter Output option.
Settings
▲
Password
Cal. purge gas
Output option
Output data
Under
Settings
menu, press
▲ and ▼ to highlight
Cal. purge gas
and press
MENU
/ .
Use
▲ and ▼ to select
on
or
off
Press
MENU
/
to store setting
Cal. purge gas
ON
Saving...
OK