Operation
64
Main menu - Options - Calibration - Configuration of calibration gas
The calibration gas is configured in this menu.
The following must be observed for this:
The calibration gas is input as a Wobbe index (Wobbe i and Wobbe s) and as a
relative density. These values are calculated from the gas components of the cali-
bration gas.
The manufacturer always uses dry gas as the basis.
Units:
xxx/m
³ and BTU/ft³ correspond to a gas temperature of 0 °C and a barometric pressure of
1013 mbar.
xxx/Sm
³ corresponds to a gas temperature of 15 °C (60 °F) and a barometric pressure of
1013 mbar.
xxx stands for MJ, kcal, or kWh.
BTU/ft³ corresponds to a gas temperature of 15 °C (60 °F) and a barometric pressure of
1013 mbar.