Transport, installation and acceptance
31
6.3.4 Process gas
DANGER
The connecting parts need to be clean and free of residue. Impurities can
enter the combustion calorimeter and cause incorrect measurements and/or
damage.
Gas inlet pressure depending on the gas between 20 and 40 mbar
table on page 8
The inlet pressure for the gas connections must not exceed the values
specified in the technical data of the combustion calorimeter.
Process gas must be free from impurities and condensate.
Each connection needs to be carefully checked for leaks. If there are any
leaks, the system will draw air, and the measurements will be incorrect.
Do not use sealing agents to seal the gas connections. Sealing agents can
distort measurements.
Only use suitable pipes.
WARNING
Only qualified trained staff must install the gas connections!
The operator must ensure to emit the exhaust gases into a safe environment!
Summary of Contents for CWD2000 EX
Page 1: ...April 2015 V1 04 Original operating instructions Combustion calorimeter CWD2000 EX ...
Page 6: ...Table of contents 6 ...
Page 9: ...Technical data 9 ...
Page 10: ......
Page 12: ...Table of contents 12 ...
Page 20: ...Table of contents 20 2 3 ...
Page 21: ...Safety devices 21 4 5 ...
Page 38: ...Table of contents 38 ...
Page 42: ...Table of contents 42 ...
Page 48: ...Table of contents 48 9 6 Menu structure ...
Page 81: ...Operation 81 ...
Page 82: ......
Page 86: ...Table of contents 86 ...
Page 90: ...Table of contents 90 ...
Page 98: ...Table of contents 98 ...
Page 100: ...Table of contents 100 ...