6.3 Installing and connecting
6.3.1 Setup site
The installation location of the combustion calorimeter must meet the following
requirements:
space of at least 50 cm to the front and below
entrance and exit as air locks
lockable windows
protected from direct weather impact and direct sunlight
stable room temperature
provide sufficient quantities of clean ambient air for undistorted
measurements
Ensure that the load-bearing capacity of the wall is sufficient
WARNING
The operator is responsible for ensuring the escape routes!
6.3.2 Ensure extractor hood remains unobstructed so that exhaust gas can
flow freely and heat cannot accumulate. Room ventilation
WARNING
Risk of injury due to the temperature of exhausts / housing!
The exhaust gas temperature is between 8 - 20°C above housing temperature!
Use PPE against burns!
NOTE
In case of adverse flow / installation conditions, appropriate deflector plates must
be installed that prevent direct flow of fresh are onto the combustion calorimeter.
6.3.3 Attachment
Further information:
enclosed documentation!
Summary of Contents for CWD2000 EX
Page 1: ...April 2015 V1 04 Original operating instructions Combustion calorimeter CWD2000 EX ...
Page 6: ...Table of contents 6 ...
Page 9: ...Technical data 9 ...
Page 10: ......
Page 12: ...Table of contents 12 ...
Page 20: ...Table of contents 20 2 3 ...
Page 21: ...Safety devices 21 4 5 ...
Page 38: ...Table of contents 38 ...
Page 42: ...Table of contents 42 ...
Page 48: ...Table of contents 48 9 6 Menu structure ...
Page 81: ...Operation 81 ...
Page 82: ......
Page 86: ...Table of contents 86 ...
Page 90: ...Table of contents 90 ...
Page 98: ...Table of contents 98 ...
Page 100: ...Table of contents 100 ...