background image

Cleveland Controls

 

DIVISION OF UNICONTROL INC. 

1111 Brookpark Rd 

Cleveland OH 44109

Tel: 

216-398-0330

Fax: 

216-398-8558

Email:

[email protected]

Web page: 

http://www.clevelandcontrols.com

Are you 

reading a FAX 

or a COPY of this 

bulletin? DOWNLOAD 

the full-color PDF ver-

sion of this and other 

literature at our 

website!

Cleveland Controls

Division of UniControl Inc.

 

LT-RSS-495-011.10

APPLICATION

The 

Series RSS-495-011 Sensing Switch 

Kit

 provides an economical and convenient 

solution for many residential and light com-
mercial HVAC applications. The kit contains 
a compact 

Series RSS-495 

switch with an 

adjustable set point range of 0.20" w.c. to 
1.0" w.c. The switch has been factory set 
to actuate on a pressure rise of 0.20" w.c. 
The switch has a bleed hole (.016") in the 
mounting pan. The individually boxed kit 
includes a P/N 28419 hex wrench (

1

16

") for 

field calibration, a P/N 28429 tube-to-tube 
reduction adapter, and two (2) P/N 27577-
037 slotted hex head mounting screws (#6 
x ⅜").

GENERAL DESCRIPTION & 

OPERATION

The 

RSS-495 

switch has a thermoplastic 

housing containing a diaphragm and snap-
acting switch. Duo-barbed sample line 
connectors on each side of the diaphragm 

accept ¼" ID or ⅜" ID slip-on flexible tubing. 

The .016" bleed hole in the mounting pan 
(low pressure side) prevents the build up of 
condensation in the  switch and sample line.

The electrical connection consists of three 
(3) male 90º quick-connect terminals. The 
snap action switch can be actuated by a 

positive or negative pressure, or by a pres

-

sure differential. 

MOUNTING (FIGURE 1)

Select a mounting location that is free from 
vibration. Mount with the diaphragm in 
any vertical plane in order to maintain the 

specified operating set point. Do not mount 

with the sample line connections directed 
upward. Surface mount via the two 0.14" 

slots in the integral foot bracket, using the 

two (2) slotted hex head mounting screws 

(#6 x ⅜") provided in the kit.

AIR SAMPLING CONNECTION 

(FIGURE 2)

Refer to 

Figure 2

 to identify the 

High inlet 

(H)

 and the 

Low inlet (L)

. Connect the 

sample lines as follows:

Positive Pressure Only:

 Connect the 

sample line to 

H

L

 remains open to the 

atmosphere. 

Negative Pressure Only: 

Connect the 

sample line to 

L

H

 remains open to  the 

atmosphere.

Two Negative Samples:

 Connect higher 

negative sample to 

L

; lower sample to 

H

.

Two Positive Samples

: Connect higher 

positive sample to 

H

; lower sample to 

L

.

One Positive and One Negative

: Connect 

positive sample to 

H

; connect negative 

sample to 

L

.

ELECTRICAL CONNECTIONS 

(FIGURE 3)

Before pressure is applied to the diaphragm, 

the switch contacts will be in the normally 
closed (

NC

) position as shown in Figure 3. 

Connect the application wiring to the corre-

sponding switch terminals, 

NC

NO

, and 

C

.

FIELD ADJUSTMENT
(FIGURE 4)

The adjustable set point range of the 

RSS-

495

 is 0.20" w.c. to 1.0" w.c., ±0.05" w.c. 

The switch leaves the factory calibrated 

at 0.20" w.c. To adjust the set point, insert 

the 

1

16

" hex wrench (provided with the kit) 

into the adjusting screw located between 
the mounting feet on the underside of 
the switch. Holding the switch so that the 

diaphragm is vertical, rotate the adjusting 

screw clockwise to the desired set point. The 

maximum number of turns is 4-½, covering 

Model

RSS-495-011

 

AIR PRESSURE SENSING SWITCH KIT WITH ADJUSTABLE SET POINT

the set point range of 0.20" w.c. to 1.0" w.c. 

Do not exceed the maximum number of 

turns. 

Each full turn represents approxi-

mately 0.18" w.c. Determine the number of 
turns required to approximately position the 
adjusting screw as follows:

(REQUIRED SET POINT – INITIAL SET POINT @ 0.20)

0.18

=  NUMBER OF TURNS CLOCKWISE

EXAMPLE:

REQUIRED SET POINT  =  0.65" w.c.

INITIAL (FACTORY) SET POINT  =  0.20" w.c.

(0.65  –  0.20) / 0.18  =  2.5 TURNS

Please note: 

(1)  Use a digital manometer and multi meter 
(such as the 

Cleveland Controls Series 

6650 Digital Manometer

) to achieve  pre-

cise set point calibration. Have the 

RSS-495 

switch as close to actual operating tem-
perature as possible during the calibration. 
The 

Cleveland Controls Series PVG-1 

Pressure/Vacuum Generator

 provides a 

handy air source for calibration.

(2) 

Warning: Never place an air 

pressure sensing switch into operation 

without confirming the set point first. An 

incorrect setting could result in a hazard-

ous condition.

Reviews: