9.6.7.GI-D6 Compatibility output
(
GI-D7 Only
)
The table below gives the relationship between pressure and output voltage.
When chose WIT
Table. 9-10
Measurement value outputs
Displayed pressure
value (Pa)
Output voltage
(V)
Displayed pressure
value (Pa)
Output voltage
(V)
9.99x10
-1
4.99
1.00x10
-4
1.10
1.00x10
-1
4.10
5.00x10
-5
0.50
5.00x10
-2
3.50
1.30x10
-5
0.13
1.00x10
-2
3.10
5.00x10
-3
2.50
1.00x10
-3
2.10
5.00x10
-4
1.50
Displayed pressure
value (Torr)
Output voltage
(V)
Displayed pressure
value (Torr)
Output voltage
(V)
9.99x10
-3
4.99
1.00x10
-6
1.10
1.00x10
-3
4.10
5.00x10
-7
0.50
5.00x10
-4
3.50
1.30x10
-7
0.13
1.00x10
-4
3.10
5.00x10
-5
2.50
1.00x10
-5
2.10
5.00x10
-6
1.50
The measurement output voltage can be converted to a pressure value using the following
equation.
P: Pressure value
(
Pa
)
V: Measured output voltage
(
V
)
E: V from which decimal point is rounded down
(
V
)
P=10(V-E)
×
10
-(5-E)
(Pa)