background image

SK00-7723-EQ-003-01 

Metal Ionization Vacuum Gauge 

GI-M2 

Ionization Vacuum Gauge    GI-D7 

Quick Manual 

Introduction 

This quick manual is for quick check of operation and display of the product. 
Please  refer  to  instruction  manual  attached  in  advance  for  detailed 
information about operation, precautions and safety for proper use. 
Available for download from ULVAC website. 
https://showcase.ulvac.co.jp/en 
This manual is for the following gauges. Serial Nos. 
GI-M2

S/N32000G

GI-D7

S/N30030G and higher 

1. Setting

1.1.  Initial Setting 

Mode 

Set value 

GI-M2 

GI-D7 

Sensor head setting 

SEn 

Not used 

SnS    1 

Sensitivity factor set value 

SEn(1) 

6.00 -2 

8.00 -2 

SETPOINT-1 

SEt-1 

1.00 -0 

5.00 -3 

SETPOINT-2 

SEt-2 

1.00 -1 

5.00 -4 

Range changeover set value 

rAn(1) 

-8

-7

Degas auto OFF set value 

dGS(1) 

01 

01 

Filament 

FIL 

FIL    1 

FIL    1 

Analog output 

rEC 

rEC    1 

rEC    1 

REC-HOLD set value 

rEC 

-8

-7

Activated filament setting 

Act 

Not used 

Not used 

Baud rate 

bPS 

9600 

9600 

1.2.  Selection of Sensor Head

GI-D7 Only

 

SnS:1 

 WIT                SnS:2 

 WIB 

1.3.  Selecting the Sensitivity Factor Function 

GI-M2 

GI-D7 

SEn:1 

N2 

N2 

SEn:2 

Ar 

Ar 

SEn:3 

Any value 

Any value 

*: N2 and the Ar sensitivity are calculated by normal sensitivity 

1.3.1.  Details of Sensitivity Set Values 

"Sensitivity" referred to here is the total sensitivity obtained by multiplying the 

sensitivity constant (K) of the ULVAC sensor head (M type) for nitrogen by the 
relative sensitivity (Srj) for various types of gases, where the sensitivity for nitrogen 
is given as 1. 

In case of M-11, intrinsic sensitivity 0.05 Pa

-1 

(normal sensitivity is 0.06 Pa

-1

N

2

: 0.05 Pa

-1

 

  Srj(N

2

)

 

Ar

  0.067 Pa

-1

   

    Srj(N

2

) × Srj 

  0.05 Pa

-1

×1.34 

1.4.  Setting of SetPoint 1/2 

GI-M2 

GI-D7 

Setiing 

range 

9.99×10

-0

 

  5.00×10

-8

 

WIT : 6.79×10

-1

 

  1.00×10

-5

 

WIB : 2.00×10

-2

 

  1.00×10

-6

 

1.5.  Selection of Range Changeover Function 

rAn:1 

 AUTO RANGE MODE

rAn:2 

 RANGE HOLD MODE 

1.6.  Setting of Range HOLD 

GI-M2 

GI-D7 

Setting range 

-0 ~ -8 

-1 ~ -7 

1.7.  Selection of DEGAS Function 

(1) dGS:1 

 MANUAL MODE

A single press on the DEGAS switch turns ON degassing and a second press
turns OFF degassing.

(2) dGS:2 

 AUTO MODE

A  single  press  on  the  DEGAS  switch  turns  ON  degassing.  It  is  automatically 
turned OFF when an arbitrarily designated time elapses. If it is desired to turn it 
OFF  before  the  time  elapses,  press  the  DEGAS  switch,  and  degassing  will  be 
turned OFF.

1.7.1.  Setting of DEGAS Function 

Settable numeric values are 01 to 99 minutes. 

1.8.  Setting of Filament 1/2 

FIL:1 

 Filament 1     

FIL:2 

 Filament 2 

1.9.Selection of Analog Function 

GI-M2 

GI-D7 

rEC:1 

Eeach-digit linear output mode 

rEC:2 

Pseudo-logarithmic mode 

rEC:3 

Range hold liniea mode 

rEC:4 

logarithmic mode 

GI-D6 compatibility output mode 

rEC:5 

Not used 

GI-TL3 compatibility output mode 

Status

 

Measurement value output voltage

 

Filament OFF 

0.00V 

Filament ON [Em.Valid OK] 

Voltage corresponding to measured pressure 

Filament ON [Em.Valid NG] 

0.00V 

When pressure protection is 

activated (Filament is OFF) 

0.00V 

1.9.1.  LIN output in each status 

P=V

×

10^-S

    (Pa) 

P: Pressure , V: Output voltage, S:Measured pressure range 

1.9.2.  GI-M2 Pseudo-LOG output 

P=10(V-E)

×

10^(E-8)

    (Pa

 

P: Pressure , V:Output voltage, E:V from which decimal point is rounded down 

1.9.3.GI-D7 Pseudo-LOG output 

P=10(V-E)

×

10^(E-7)

    (Pa

 

P: Pressure, V:Output voltage, E:V from which decimal point is rounded down 

1.9.4.  Range Hold Linear Output (REC-HOLD) 

(by E-4) 

1.9.5.  Logarithmic Output (GI-M2) 

P=10^(V-8) 

  V=logP+8

 

P: Pressure, V: Output voltage 

1.9.6.GI-D6 Compatibility output setting WIT 

GI-D7 Only

 

P=10

×

(V-E)

×

10^(E-5)

    (Pa) 

P: Pressure, V: Output voltage, E: V from which decimal point is rounded down 

1.9.7.GI-D6 Compatibility output setting WIB 

GI-D7 Only

 

P=10

×

(V-E)

×

10^(E-6)

    (Pa) 

P: Pressure, V: Output voltage, E: V from which decimal point is rounded down 

1.9.8.GI-TL3 Compatibility output 

GI-D7 Only

 

P=10^(V/2-5)

    (Pa) 

P: Pressure, V: Output voltage 

2. Nomenclature and Functions of Components

Name (Inscription) 

Function 

  POWER 

Pressing the top of the POWER switch turns on power 

and pressing the bottom turns it off.    When the power is 
on, the LED in the switch is lit. 

  FIL 

Used to turn on/off the filament.    The first press on the 

key turns on the filament and a second press turns it off. 

  DEG 

Used to turn on/off degassing.    The first press on the key 

turns on degassing and a second press turns it off. 
Degassing is automatically turned off after a preset time 
when used in the auto OFF mode. 

  F1 

Not used 

  O PRG 

Switch to change over to the setting mode. 

Holding down the switch for 2 seconds or more changes 
over the measurement mode/setting mode. 

 

 

Each press on the switch increments a value by one. 

 

Shifts a value to the right. 

 

Used to store a set value in memory. 

A set value cannot be changed unless this switch is 
pressed because it is not stored in EEPROM. 

Name 

Function 

 

SET1 lamp 

Lit when set point 1 is activated. 

 

SET2 lamp 

Lit when set point 2 is activated. 

 

PROTECT lamp 

Lit when filament protection is activated by an external 
input signal. 

 

REMOTE lamp 

Lit when the gauge is set in the REMOTE status by 
external control (external I/O, RS-232C). 

 

Em-Lo lamp 

GI-M2: Lit when emission current is 10 

A. 

GI-D7: Not used. 

 

Em-Hi lamp 

GI-M2: Lit when emission current is 1 mA. 

GI-N7: Not used. 

0

1

2

3

4

5

6

7

8

9

10

1.0E-05

1.0E-04

1.0E-03

O

ut

put

 V

ol

ta

ge

 (V

)

Pressure (Pa)

Pressure increase

Pressure decrease

1.0E-(N)

1.0E-(N-1)

1.0E-(N-2)

0

1

2

3

4

5

6

7

8

9

10

1.0E-08 1.0E-07 1.0E-06 1.0E-05 1.0E-04 1.0E-03 1.0E-02 1.0E-01 1.0E+00 1.0E+01

O

u

tp

u

V

ol

tage

 (V

)

Pressure (Pa)

0

1

2

3

4

5

6

7

8

1.0E-06

1.0E-05

1.0E-04

1.0E-03

1.0E-02

1.0E-01

1.0E+00

O

u

tp

u

V

ol

tage

 (V

)

Pressure (Pa)

0.01

0.1

1

10

1.0E-07

1.0E-06

1.0E-05

1.0E-04

1.0E-03

1.0E-02

1.0E-01

1.0E+00

O

u

tp

u

V

ol

tage

 (V

)

Pressure (Pa)

0

1

2

3

4

5

6

7

8

9

10

1.0E-08 1.0E-07 1.0E-06 1.0E-05 1.0E-04 1.0E-03 1.0E-02 1.0E-01 1.0E+00 1.0E+01

O

u

tp

u

V

ol

tage

 (V

)

Pressure (Pa)

0

1

2

3

4

5

6

1.0E-05

1.0E-04

1.0E-03

1.0E-02

1.0E-01

1.0E+00

O

u

tp

u

V

ol

tage

 (V

)

Pressure (Pa)

0

1

2

3

4

5

6

1.0E-06

1.0E-05

1.0E-04

1.0E-03

1.0E-02

1.0E-01

1.0E+00

O

u

tp

u

V

ol

tage

 (V

)

Pressure (Pa)

0

1

2

3

4

5

6

7

8

9

10

1.0E-06

1.0E-05

1.0E-04

1.0E-03

1.0E-02

1.0E-01

1.0E+00

O

u

tp

u

V

ol

ta

ge

 (

V

)

Pressure (Pa)

1

2

3

4

5

6

7

8

1

2

3

4

5

6

7

8

9

10

Pa