SK00-9663-EI-003-00
39
3)
SP1: Measured value output in each state
With the SP1 setting, the measured pressure is output as a DC 0V to 10V signal.
(I/O connector: 8pin [pressure signal ] -15pin [GND])
The table below shows the measured value output in some conditions that can occur
during measurement.
Operating state
Measured value output
Normal measurement
0.1V to 5.0V
3000Pa or higher
5.0V or higher
Below measurable lower limit
0.1V or less
Sensor error such as filament
disconnection
9.0V or higher
4seconds after startup
5.4V or less
Failure of power supply, sensor error,etc.
DC 0.5V or less
※
The set point function will be disabled for 4seconds after startup.
※
For signals other than measured value output, refer to Sections 5.2 and 5.3 of this
manual.
9.3.
Other compatibility
It also supports the measured value output corresponding to the conversion formula to
pressure shown below. However, the basic specifications other than those described in
this section are the basic specifications of SW100 shown in Section 1.2, and the sensor
can also use the SWP series dedicated to SW100.
PSG compatibility
1)
PSG: Compatible specifications
By setting the OUTPUT of SW100-A to PSG with UL-MOBI, the measured value output voltage
corresponding to the conversion formula shown below is output.
・
Measured value Output voltage value-Pressure conversion formula
P=10^((V-3.572)/1.286) (P: Pressure[Pa]
、
V: Measured output voltage[V])
・
Measurement pressure range
5×10
-2
Pa to 1×10
+5
Pa
As the sensor, only the SWP series dedicated to SW100-A can be used.
2)
PSG: I/O specifications
SW100-A: By setting OUTPUT to PSG with UL-MOBI, it can be used with the measured output
voltage corresponding to the above conversion formula.
For pin layout, refer to Section 9.4 of this manual.
SW100-R: Not compatible.
※
Depending on the internal circuit structure of the replacement source vacuum
gauge, set point output may not be possible.
※
The I/O connector is a D-sub15pin connector (male, M2.6mm screw)
Summary of Contents for G-TRAN SW100-A
Page 7: ...SK00 9663 EI 003 00 VII Revision history Date Number Reason 2022 10 3 00 First edition ...
Page 69: ...SK00 9663 EI 003 00 69 15 EC DECLARATION OF CONFORMITY ...
Page 70: ...SK00 9663 EI 003 00 70 16 UK DECLARATION OF CONFORMITY ...
Page 71: ...SK00 9663 EI 003 00 71 17 Certificate of decontamination ...
Page 72: ...SK00 9663 EI 003 00 72 18 Drawings 18 1 SW100 A R dimensions ...
Page 74: ...SK00 9663 EI 003 00 74 18 3 Display unit cable L ...