22.3.
Standard Data Format
The table below gives the standard data sending/receiving format.
:
AD0
AD1
CMD
D0
・・・・・・
Dn
SH
SL
CHKH CHKL
CR
:
Colon
AD0
Device address high order (0 – 9)
AD1
Device address low order (0 – 9)
CMD
Various commands (note uppercase character/lowercase character)
D0
Data
Dn
Data
SH
High order of status
SL
Low order of status
CHKH High order of checksum (0 - 9, A – F)
CHKL Low order of checksum (0 - 9, A – F)
CR
Carriage return
* A command consists of uppercase characters and lowercase characters of alphanumeric
characters.
* Checksum is an exclusive logic sum (XOR) from AD0 to SL.
* Convert all with the ASCII code.
22.3.1.
Command list
Command
Description
Command
Description
‘D’
Reads measurement value and status
‘1W’
Writes setpoint 1
‘SR’
Reads status
‘2W’
Writes setpoint 2
‘SW’
Writes status
‘3W’
Writes setpoint 3
‘T’
Reads software
‘ZER’
Adjusts SW1 zero point
‘1R’
Reads setpoint 1
‘ATM’
Adjusts SW1 atmospheric pressure
‘2R’
Reads setpoint 2
‘CLR’
Resets adjustment of SW1 zero
point, atmospheric pressure
adjustment
‘3R’
Reads setpoint 3
‘CZR’
CCM zero point adjustment
‘CCR’
Resets CCM zero point adjustment
22.3.2.
When received normally
The following is returned when data is normally received.
When a pressure value is returned
:
AD0
AD1
D
X
.
X
X
E
±
X
X
SH
SL
CHKH
CHKL
CR
When writing setpoint or adjusting
:
AD0
AD1
o
CHKH
CHKL
CR
* ‘o’ is a lowercase character.
22.3.3.
When receiving is faulty
The following is returned when not normally received or when there is no command.
:
AD0
AD1
n
CHKH
CHKL
CR
* ‘o’ is a lowercase character.