Operating state
Analog output voltage
During normal measurements
Voltage corresponding to the measured pressure
0.27 to 8.75 V
1x10
+4
Pa or higher
8.75 V
SH2 gauge FIL OFF
Voltage corresponding to the measured by SPU
5 V to 8.75V
SH2 error
(Errors such as a filament break)
Voltage corresponding to the measured by SPU
5 V to 8.75V
SPU error
(Errors such as a filament break)
9.9 V or higher
Power supply voltage abnormality,
sensor unit fault, etc.
0.1 V or less
14.6.2.
Filament OFF signal
Signal that turns off the filament. To turn off the filament, connect it to GND with the external
switch. In the case of front panel operation or RS485 communication setting, input of this signal
will be invalid.
14.6.3.
Degas ON signal
Signal that turns on degassing. To turn it on, connect it to GND with the external switch. In the
case of front panel operation or RS485 communication setting, input of this signal will be invalid.