49
13.6.5.
Pressure protection signal
If the sensor unit exceeds the higher limit of measurement pressure, pressure protection will be
actuated. If it is actuated, the open collector will be Lo output.
Photocoupler rating: 30V
MAX
, 50mA
MAX
, 70 mW
For more information about pressure protection and actions in case of pressure protection signal
being turned ON, refer to the sensor unit manual.
13.6.6.
Emission valid signal
When the emission current is flowing normally or not is output as a signal. If emission current
is flowing normally, the open collector will be Lo output.
Photocoupler rating: 30V
MAX
, 50mA
MAX
, 70 mW
For more information about the emission valid signal and corrective actions in case when the
emission valid signal is turned off, refer to the sensor unit manual.
13.6.7.
Setpoint
For how to adjust and use the setpoint, refer to Section 7.
Photocoupler rating: 30V
MAX
, 50mA
MAX
, 70 mW
13.6.8.
RS485 communication
For RS485 communication, refer to Section 22.