6.2 Connection of GN-10i to power source
(1) Use tools such as circuit tester to check that the supply voltage is within the
appropriate range as described in “5. Utilities”.
(2) Plug the power-supply cable which comes out from the back of GN-10i
(3) Turn on the power breaker on the back of the GN-10i.
6.3 Dry air connection to IM-120
(1) Attach the connector for tubes to the dry air valve. Screw size is Rc1/4.
(2) Onto the connector attached as in the step (1), attach the tube for dry air supply.
Connect the other side of the tube to "AIR IN" connector on the back side of
IM-120.
6.4 Utility connection of EMP-14A
Remove the back panel of EMP-14A and locate holes for utility connection on the
bottom. Connect cables through these holes including power-supply, Nitrogen
tube and remote cables of Nitrogen generation system, signal cables for automatic
transfer destination.
Ensure that main power source is disconnected before making the connection to
power-supply.
WARNING
Following procedure must be undertaken by an authorized electrician familiar with the
structure and risk of the product. Failure to observe this warning may result in severe
bodily injury or fatal accident.
WARNING