C30VR Compressor Installation, Operation, Maintenance and Troubleshooting
8-4
ULVAC CRYOGENICS INCORPORATED
8.3 Alarm code descriptions
<
Alarm status
>
→
lit up
→
off
Reverse Phase, Open Phase, Low Volt
(
Reverse phase, open Phase, or low voltage
)
When the state of the compressor main power supply corresponds to following
either and the state continues 3 seconds or more, all LED indicators light up.
1.
Reverse phase
2.
Open phase
3.
Low voltage
(
in case of nominal 170V or less
)
When a voltage drop arises at the time of starting, a compressor stops immediately,
however the alarm is not turned on. It is because the voltage value returns to
normal, shortly after stopping. In this case, a start and a stop might be repeated.
Check and make sure that the line voltages (R-S, S-T, and T-R) of the power cable
are within the rated value. If these voltages are within the rated value and all
indicators are light up, there is a possibility of reverse phase. Check the wiring of
the main power supply referring to “Section 5.2
(
3
)
Checking the compressor
status and main power phase sequence”.
If the input power supply voltage is low, check the input power line and the user’s
power supply that has distributed electric power to the compressor.
When there is an open phase(s), all indicators should be blinked or turned off.
TS2
Over Temperature
(
Over temperature
)
TS2 monitors the temperature of the compressor pump. If TS2 is activated it
means the compressor pump is in an unrecoverable abnormal condition. Please
contact our Service Engineering Division or the nearest customer support center
immediately.
1
2
3
4
5
6
7
8
1
2
3
4
5
6
7
8
STS
Troublesh
cedures described above must be undertaken by an authorized
electrici
mechanical personnel familiar with the structures and the risks of the
compress
s related components. To ensure safe, reliable system performance, read
this ma
as other relevant manuals completely to gain thorough understanding
before be
rk.
When
ssor does not need power supply to follow the procedure, shut down the
power of the compressor and lock it in the OFF status. Failure to observe this caution may
result in severe bodily injury or fatal accident.
CAUTION
ooting pro
an or
or and it
nual as well
g
ing wo
a compre
inn
Summary of Contents for C30VR Series
Page 2: ......
Page 8: ...F 6 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Page 16: ...S 4 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Page 18: ...IW 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Page 108: ...CS 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Page 110: ...Revision History R 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...