![Ultrawave QS13 Series Instructions Manual Download Page 2](http://html1.mh-extra.com/html/ultrawave/qs13-series/qs13-series_instructions-manual_837647002.webp)
QSx Operating Manual
2
C6054101 (June 2020)
Important Information
Please read these instructions carefully before using this
QSx
Ultrasonic
Bath. In particular, read and understand all warnings, cautions, warning
labels and safety instructions.
Should there be any comments regarding the equipment or these
instructions, please contact Ultrawave Ltd
Scope
The scope of this Operator Instruction Manual is for the installation and
safe use of the
QSx
ultrasonic bath.
Electrical Safety instructions
The QSx Ultrasonic Bath is designed to be connected to a 230V 50Hz
electrical supply via a three pin BS1363 plug.
Always connect to a fully earthed supply using the moulded plugs provided.
It can be dangerous to operate an ultrasonic bath without an earth
connection.
The unit is supplied with a removeable power lead rated at 5A. Do not use
a power lead rated less than 5A.
The removeable power lead is the disconnect device and should remain
accessible while the unit is in use.
The IEC C14 electrical socket on the rear of the unit is fitted with 2x T5A
(slow blow) fuses. Never fit a fuse of a higher rating.
The unit is rated at 500W.
Ensure that any excess power cable is stored neatly.
Using this machine in a manner not specified by Ultrawave may lead to
damage or electrical safety impairment.
Ultrawave will not be responsible for damage or injury caused by incorrect
use of the equipment.