SonixTouch Q+ Service Manual
00.053.204, Revision A
Chapter 7: Field Service Components
7-29
7.4.3
Removing the Front Block/Ultrasound Module
To remove the front block/ultrasound module without removing the modulo, follow the instructions in
before continuing to the instructions below.
To remove the front block/ultrasound module after removing the modulo, follow the instructions in
and
before continuing to the instructions below.
Equipment/Tools Required:
• #2 Phillips screwdriver.
To Remove the Front Block/Ultrasound Module:
1.
With the Front Block/Ultrasound Module partially open, locate the cable tie securing the PCI cables to the
inside of the door. Reach in and
carefully
cut the cable tie securing the ribbon cables to the interior of the
modulo door.
2.
Once the cable tie is removed, open the Front Block/Ultrasound Module wider and disconnect the five attached
cables.
Warning:
Do not perform any internal system maintenance if the UPS breakers are turned on.
Caution:
Always wear a grounding strap when opening and working inside the modulo.
Note:
To avoid confusion during reinstallation label the cables as each one is detached.
1
3
2
5
4