SonixTouch Q+ Service Manual
00.053.204, Revision A
Chapter 3: System Installation
3-17
3.3.4
UPS Configuration (If Applicable)
Upon delivery, the main (system) power switch and UPS battery breakers will be turned off. The battery itself may
be completely drained of power. During installation, the technician must ensure the system is left plugged in,
allowing the battery to charge. A full charge will take approximately 6.5 hours.
Service (and emergency) access to the UPS breakers is provided through two holes near the base of the left side
shroud (side opposite to the transducer connectors).
Figure 3-6: UPS Breaker Access
Table 3-3: Breaker Types and Locations
To Turn a Breaker ON/OFF Without Removing the Side Shroud:
1.
Using the openings provided to access the UPS breakers, press the bottom of the breaker to turn it off.
2.
Press the top of the breaker to turn it on.
1
Battery Breaker
2
AC Input Breaker
3
System front
Warnings:
NEVER
let liquid from any source enter the UPS. Failure to do this may result in accidental
shorts, shocks or
electrocutions
.
DO NOT attempt to service this product yourself.
Attempting to open the UPS may cause exposure to lethal
voltages within the unit even when it is apparently not operating and the input wiring is disconnected from the
electrical source. Should the UPS require maintenance or replacement, only qualified Ultrasonix Service
Technicians may perform service as detailed in the Service Manual.
Use only the UPS battery recommended and supplied by Ultrasonix Medical Corporation.
For UPS and battery service issues, contact Ultrasonix Technical Support.
If the battery is removed from the system, it is the responsibility of the customer to dispose of it in accordance with
all local regulations and laws.
Note:
When turned off, there is a visible white marking at the top of the breaker.
1
1
2
2
3
Summary of Contents for SonixTouch Q+
Page 1: ...Analogic Ultrasound Service Manual SonixTouch Q Ultrasound System...