Verify the system is to be operated in a room that meets the electrical requirements listed below.
Table 3-4: Electrical Specifications
3.1.2.1 Instrument Input Power Rating
An instrument input power rating has been established to clarify the maximum power requirement of the system
with all accessories. Depending upon the accessories and options installed, the system may not draw the full
amount of power listed below:
Table 3-5: Instrument Power Input Rating (with optional UPS)
Table 3-6: Instrument Power Input Rating (without optional UPS)
For optimal system performance, use a dedicated, interference-free, isolated, grounded wall outlet. To ensure
grounding reliability, use a hospital-grade power cord and connect it only to an equivalent hospital-grade socket.
The specifications of the hospital-grade power cord as follows:
Table 3-7: Hospital-Grade Power Cord Specifications
Electrical Rating
Value
Measurement (System Only)
100V—120V @ 50/60 Hz
200V—240V @ 50/60 Hz
Current Draw (System Only)
120 VAC @ 7.0A
240 VAC @ 4.0A
Power Draw (System Only @ peak)
960 VA
Leakage Current (System Only)
< 100 uA
Note:
Refer to OEM peripheral manuals for peripheral-specific electrical ratings.
System Voltage, VAC
Nominal System Power Usage,
including isolated accessory power
System Power Usage
Isolated Accessory
Power Available
120V
7.0A Max
5.2A Max
1.8A Max
240V
4.0A Max
3.25A Max @120V
0.75A Max @ 120V
System Voltage, VAC
Nominal System Power Usage,
including isolated accessory power
System Power Usage
Isolated Accessory
Power Available
120V
7.0A Max
5.2A Max
1.8A Max
240V
4.0A Max
3.25A Max @120V
0.75A Max @ 120V
Input Voltage
Hospital Grade Power Cord Specifications
100-120V ~, 50/60Hz
125Vac, 15A, 3 wire, 18 AWG, grounding type, 5-15P Hospital Grade plug cap, less than 6 m long,
CSA & UL approved
200-240V~, 50/60Hz
250Vac, 15A, 3 wire, 18 AWG, grounding type, 6-15P Hospital Grade plug cap, less than 6 m long,
CSA & UL approved
Summary of Contents for SonixTouch Q+
Page 1: ...Analogic Ultrasound Service Manual SonixTouch Q Ultrasound System...
Page 2: ......
Page 4: ......
Page 14: ...Chapter 2 System Specifications 00 053 204 Revision A SonixTouch Q Service Manual 2 2...
Page 42: ...Chapter 4 Performance Testing 00 053 204 Revision A SonixTouch Q Service Manual 4 4...
Page 120: ...Chapter 8 DICOM 00 053 204 Revision A SonixTouch Q Service Manual 8 24...
Page 147: ......