Appendix A: Troubleshooting Solutions
00.053.129, Revision C
SonixTablet Service Manual
A-4
A.5
IMAGE HAS THREE (3) UNIFORM BLACK AREAS (LEFT, MIDDLE, RIGHT)
A.5.1
Test US Module
To Test the US Module:
1.
Disconnect all transducers.
2.
Connect a completely different transducer.
3.
Restart the system.
4.
If this fails to solve the problem, contact Ultrasonix Technical Support.
A.6
REGEDIT – ENSURE SETTINGS ARE CORRECT
To Edit REGEDIT:
1.
Ensure a USB keyboard is connected to the system.
2.
Press the
WINDOWS START KEY
and select.
1.
Type
regedit
and press
Enter
.
2.
In the registry go to
HKEY_CURRENT_USER > Software > Ultrasonix > Exam > Settings
.
3.
Make the changes advised by the Ultrasonix Technical Support Representative.
Note:
Contact an Ultrasonix Technical Support Representative for details on the required changes.
Summary of Contents for SonixTablet
Page 1: ...SonixTablet Ultrasound System Service Manual...
Page 2: ......
Page 4: ......
Page 10: ...List of Service Drawings 00 053 129 Revision C SonixTablet Service Manual ii...
Page 30: ...Chapter 3 System Installation 00 053 129 Revision C SonixTablet Service Manual 3 14...
Page 34: ...Chapter 4 Performance Testing 00 053 129 Revision C SonixTablet Service Manual 4 4...
Page 38: ...Chapter 5 Software 00 053 129 Revision C SonixTablet Service Manual 5 4...
Page 44: ...Chapter 6 Peripherals and Accessories 00 053 129 Revision C SonixTablet Service Manual 6 6...
Page 84: ...Chapter 9 Network Configuration 00 053 129 Revision C SonixTablet Service Manual 9 8...
Page 100: ...Chapter 11 Maintenance 00 053 129 Revision C SonixTablet Service Manual 11 14...
Page 118: ...Appendix C Ultrasonix Limited Warranty 00 053 129 Revision C SonixTablet Service Manual C 2...
Page 120: ...Appendix D Service Drawings 00 053 129 Revision C SonixTablet Service Manual D 2...