Ultradent Products Ultrapro Tx Air Maintenance And Troubleshooting Manual Download Page 1

Thank you for choosing the Ultrapro Tx Air Hygiene Handpiece from Ultradent

When properly cared for, your motor will provide years of quality service. Please follow these guidelines to ensure proper functioning over 
the life of your equipment.

Oil and Lubrication 

Lubrication of the motor and nose cone is necessary to keep all the moving parts running properly. Oil also helps flush debris out of the 
motor, helping to prevent buildup or wear and tear. 

Preventing Buildup 

Motors coated with debris may not work consistently or may lose power when pressure is applied to the motor. 

*For complete sterilization and maintenance instructions, please refer to the product IFU.

Ultrapro

®

 Tx Air Handpiece

  

Maintenance and Troubleshooting Guide

OIL

OIL

Figure 2

Lubricating Motor

Lubricating Nosecone      

OIL

Figure 1

OIL

OIL

Figure 2

Lubricating Motor

Lubricating Nosecone      

OIL

Figure 1

OIL

OIL

Figure 2

Lubricating Motor

Lubricating Nosecone      

OIL

Figure 1

After sterilizing, place two drops 
of handpiece oil into the motor 
air tube (Fig. 1) and 1–2 drops 
of oil into both ends of the 
nosecone (Fig. 2). To properly 
disperse the oil, run the motor 
for several seconds. 

Apply one drop of oil to the O-ring 
once per week to keep the nose 
cone swivelling smoothly. 

To prevent debris inside the motor: 

1.

  Have your air compressor serviced regularly                                     

2.

  Clean or replace air hoses regularly 

Always check that your compressed air is filtered, dry, and free of oil.

U LT R A D E N T. C O M

800.552.5512

Poorly maintained motor

Well-maintained motor

© 2018 Ultradent Products, Inc. All Rights Reserved. 062918