ULTIMA 250 SERIES

 

Owners Manual 

 

Summary of Contents for 250 Series

Page 1: ...www UltimaDentalSystems com 888 900 8584 ULTIMA 250 SERIES Owners Manual...

Page 2: ...humb screw located at the top of the unit The two suction feet provided are to ensure more stability for the unit on a table top surface even if the water bottles are empty To make a more permanent in...

Page 3: ...a wall beam Use a stud finder which can be found in most hardware stores to locate a beam in the area the Ultima 250 will be installed However if this is not possible a second option is to use Drywall...

Page 4: ...re attach the other half tighten the screws on the pole mount until tight and secure Both Table Top and Arm units include a standard 12 arm attachment Additional 6 8 12 and 16 arm attachments are als...

Page 5: ...female quick disconnect fitting on the side of the unit labeled Filtered Air In Set the air pressure from the compressor to 70 Psi NEVER EXCEED 80 Psi GOING INTO THE UNIT If you are using a compresso...

Page 6: ...the male quick disconnect fitting with the shut off valve to the female disconnect fitting on the side of the unit labeled Filtered Air In Set the air pressure from the nitrogen source to 70 Psi NEVER...

Page 7: ...ISTILLED WATER AFTER USE FOR THIRTY SECONDS TO A MINUTE TO NOT ALLOW THE CHOLHEXIDINE TO AND FORM PARTICLES INSIDE THE TUBING LINES WHICH CAN CAUSE CLOGS AND BLOCKAGE WITHIN THE UNIT 4 The white cap n...

Page 8: ...he holes are small adjustment screws To decrease individual air line pressure turn adjustment screw clockwise To increase individual air line pressure turn counter clockwise To install hand pieces sim...

Page 9: ...e very little maintenance 1 Always keep handpieces oiled with liquid handpiece oil 2 Disconnect air source going into the unit each night to allow the unit to vent 3 Always remember to flush out the t...

Page 10: ...transducer The transducer consists of ceramic disks stacked next to one another These ceramic disks have high quartz content and when excited with electricity they compress and decompress causing a co...

Page 11: ...oesn t heat up so only the tip requires cooling reducing the amount of water used tremendously Adjusted the Water Flow 1 Screw on the 1 tip Universal onto the handpiece until it is hand tight Then use...

Page 12: ...off water when using a tip without a water port Handpiece and tubing If using irrigating fluids other than water the cord and handpiece should be flushed out with distilled water for at least one minu...

Page 13: ...Handpiece damaged 4 Electrical contacts on cord are loose 5 Tip worn or distorted 1 Unplug the scaler unit and wait 30 seconds to reset the scaler module 2 Replace tip worn tips will not allow the sca...

Page 14: ...ized the bottles should be rock hard If no water continues to be the issue the problem may be more involved and may require a professional dental equipment technician Other causes of no water issues c...

Page 15: ...properly onto the polisher nose cone 1 Faulty or damaged nose cone chuck 2 Prophy angle stem has broken inside of the nose 1 The nose cone chuck will need to be repaired or replaced Contact Ultima Den...

Page 16: ...side down for thirty seconds to allow the oil to travel through the polisher It is important that the polisher be run upside to avoid the thin handpiece oil from sliding back down the handpiece tubing...

Page 17: ...drill is most likely damaged and needs to be replaced 1 Remove the end cap located on the back of the drill Check for debris this is a sign that parts inside the turbine such as the bearings may have...

Page 18: ...of the two larger barbs located on the bottom of the drill Re attach the hand piece and run the drill upside down for thirty second to a minute to allow the oil to circulate through the drill Turning...

Page 19: ...ent spare parts for T64 highspeed drills are still available for purchase If you have any questions or need assistance ordering parts for T64 highspeed drills contact Ultima Dental Systems at 888 900...

Page 20: ...warranty if defective out of the box and damaged during shipping All parts on Ultima compressors are covered for 5 years if defective The Lowspeed Polisher is covered under warranty for 1 year if defe...