Ulthera, Inc.
Page
29
4.5.3.
Recommended Separation Distances between Portable and
Mobile RF Communications Equipment and the Ulthera System
Ulthera is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The Ulthera user can help prevent
electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the
Ulthera system as recommended below, according to the maximum output
power of the communications equipment.
Table 4.8
Recommended Separation Distances
RATED
MAXIMUM
OUTPUT
POWER OF
TRANSMITTER
IN W
SEPARATION DISTANCE ACCORDING TO FREQUENCY OF TRANSMITTER,
METERS
150 kHz to 80 MHz
d=1.17
√
P
80 MHz to 800 MHz
d=1.17
√
P
800 MHz to 2.5 GHz
d=2.33
√
P
0.01
0.12 m
0.12 m
0.23 m
0.1
0.37 m
0.37 m
0.74 m
1
1.2 m
1.2 m
2.3 m
10
3.7 m
3.7 m
7.4 m
100
12 m
12 m
23 m
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
4.6.
Disposal
Depleted transducers should be disposed of in accordance with federal,
state, and local regulations.
Summary of Contents for DeepSEE DS 10 -1.5
Page 2: ...Page 2 Ulthera Inc THIS PAGE INTENTIONALLY LEFTBLANK...
Page 38: ...Page 38 Ulthera Inc 6 1 1 Face Neck Amplify...
Page 39: ...Ulthera Inc Page 39 6 1 2 General Regions Face Neck...
Page 69: ...Ulthera Inc Page 69 THIS PAGE INTENTIONALLY LEFTBLANK...
Page 70: ...Page 70 Ulthera Inc THIS PAGE INTENTIONALLY LEFTBLANK...
Page 71: ...Ulthera Inc Page 71 THIS PAGE INTENTIONALLY LEFTBLANK...