Instruction for use (rev 141119)
e.ON 200 NEO / e.ON 250 NEO
UGIN DENTAIRE
29
6.
MAINTENANCE AND TROUBLE-SHOOTING
Before to carry out the maintenance work and/or cleaning,
e.ON 200/250 NEO has to:
- disconnected from the main power supply
(switch OFF and unplug to mains socket)
- at room temperature: wait the cooling after use (risk of burns and/or damage)
6.1.
MAINTENANCE
Observe the following recommendations to ensure the longevity of the device:
-
Do not use abrasive, solvents or flammables products for cleaning (outside and inside)
-
Clean the machine outside with a soft, dry cloth
-
Remove any traces on the screen with a soft cloth and a small amount of special glass cleaner. Cleaning
wipes for optic products can also be used
-
Clean the joint of the tray with alcohol regularly, to ensure the proper functioning of the vacuum
system. To avoid any risk of adhesion, it is recommended to sprinkle it with a thin layer of talc powder
-
e.ON 250 NEO: disconnect the compressed air inlet of the furnace about once a month
This allows to automatically purge the compressed air inlet filter of the furnace to evacuate any
moisture in the unit
-
Visual inspection of the support pressing and the firing base before use. If necessary, clean with a
vacuum cleaner to remove any particles (dust). Do not use compressed air. If the support/base is
damaged or cracked: do not use and replace-it
-
Use a vacuum cleaner to remove any dust in the chamber firing or in the tray – do not use
compressed air