![UFSK 500 ECO User And Service Manual Download Page 41](http://html1.mh-extra.com/html/ufsk/500-eco/500-eco_user-and-service-manual_3240233041.webp)
500 ECO
Manual_500 ECO_SN0001-following_0520
www.ufsk-osys.com
User and Service Manual
Page 41 of 50
9.2. Electrotechnical Data
Power supply
Pri. Volt: 100-240V, 50/60 Hz Ac; Sec. Volt: 29V DC
Rating
650 VA at 240 Vac / 440 VA at 100 Vac
Duty cycle/pause cycle of
the lifting column and drives
10% max.; 2 minutes max. of continuous duty
followed by a 18 min brake
Protection class of the drives IP54
Safety functions
Overload protection
Overcurrent trip
Electrical design
DIN EN 60 601-1, protection class II, type B,
degree of protection IP24
9.3. Classification
Medical device
CE conformity
In the EU:
Class I, without measuring function, non-sterile;
Conformity assessment procedure in accordance
with Annexes II and III, Regulation (EU) 2017/745, Art.52 (7)
in the US/FDA:
Class I, Conformity assessment procedure
in accordance with 21 CFR Part 886, 510(k) Exempt
Degree of Protection
according to DIN EN 60 601-1
Protection class II, Type B
Degree of Protection of Enclosure
according to DIN EN 60529
Degree of Protection IP54
Technical Safety Inspection
according to DIN EN 62353
Once per year
9.4. Ambient conditions
Ambient conditions
for operation
Temperature: +10°C…+40°C
Relative ambient humidity: 30%...75%
Air pressure 700hPa…1,060hPa
Ambient conditions
for shipping and storage
in shipping packaging
Temperature: -20°C…+50°C
Relative ambient humidity: 10%...95%
Air pressure: 660hPa…1,080hPa
In the interest of technical developments we reserve the right to make modifications in design and scope
of supply without prior notice.