background image

DMG150

mbar   Pa  mm-Hg

Microns mTorr        /sec

Set Leak Rate Temp Run

MODE

MENU

DMG150

Digital Micron Gauge

1-800-547-5740

www.ueitest.com

 • email: 

[email protected]

INSTRUCTION MANUAL

ENGLISH

IMPORTANT SAFETY WARNINGS

 

 

WARNING

Read entire Safety Notes section regarding potential 

hazard and proper instructions before using this meter. 

In this manual the word “

WARNING

” is used to indicate 

conditions or actions that may pose physical hazards 

to the user. The word “

CAUTION

” is used to indicate 

conditions or actions that may damage this instrument. 

 

 

WARNING

To ensure safe operation and service of the tester, follow 

these instructions. Failure to observe these warnings can 

result in severe injury or death.

 

 

WARNING

This gauge is designed for trade professionals who 

are familiar with the hazards of their trade. Observe all 

recommended safety procedures that include proper use 

of personal protective equipment that includes safety 
glasses and gloves.

FEATURES

•  Measures in microns, pascals, Millibar, mmHg, mTorr
•  Range 0 to 15,000 microns
•  Maximum over pressure 700 psi
•  Instant warm-up time
•  Instant response time
•  Backlit display
•  Compact and light weight
•  Oil contamination detection
•  Sleep mode
•  Ambient temperature measurement
•  5 units of measurement
•  Real time leak measurement
•  Time period leak measurement
•  Sensor fail detection

GENERAL SPECIFICATIONS

•   Operating Temperature: 

10˚ to 122˚F (-12˚ to 50˚C)

•   Backlight: 

Yes

• Over-range:

 “OL” is displayed 

• Dimensions: 

4.3’ X 1.3’ X 1’

•  Item Weight: 

2.7 oz

• Calibration: 

Recommended annually 

 

 

• Accuracy: 

0 to 9,999 ±5%rdg ±5 microns 

   

10,000 to 15,000 ±10% rdg

OVERVIEW

DMG150

mbar   Pa  mm-Hg

Microns mTorr        /sec

Set Leak Rate Temp Run

MODE

MENU

D. 

Menu/Check 

Button

E. 

Change/X  

Button

F. 

Dust Cap

A. 

LCD 

Display

B. 

Power/

Backlight 

Button

C. 

Mode 

Button

G. 

Battery 

Cover

H. 

Vacuum Port

J. 

Dust 

Cap Removed

I. 

Serial Number

OPERATING INSTRUCTIONS

Basic Operation Mode

1.  Press Power button to power on the gauge.
2.  Press 

Mode Button

 to change the operation to Vacuum  

 

  Reading, Temperature Reading, Leak Rate Test, Leak Test or    

  Run mode.

Vacuum Reading Mode

1.  The Vacuum reading is shown on the display with Vacuum    

 unit. 

 

 CAUTION: 

OL

” will be shown if maximum Micron or    

  psi limit is exceeded (

15,000 Micron/700 psi

 or unit reading   

  of equal value).
2.  User may press 

Mode Button

 to skip Vacuum Reading mode   

  or enter Temperature Reading mode.
3.  In Vacuum Reading mode, press Menu Button or Change  

 

  Button to select the unit to mBar, Pa, mmHg, mTorr or Micron. 
  The Vacuum indicator will be shown to indicate the Vacuum    

  unit selected.

Temperature Reading Mode

1.  The Vacuum reading is shown on the display with  

 

  Temperature measurement unit. 
2.  User may press 

Mode Button

 to skip Temperature Reading    

  mode or enter Leak Rate Test mode.
3.  In Temperature Reading mode, press 

Menu/Check Button

    

 or 

Change/X Button

 to select temperature measurement    

  unit to ˚F or ˚C

Leak Rate Test Mode

1.  In Leak Rate Test mode, the leak rate test time base is 1  

 

 second.
2.  The indicator “

Leak Rate

” and “

/sec

” will be shown. The    

  real time leak rate is updated on the display once per second.
3.  User may press 

Mode Button

 to skip Leak Rate Test mode    

  and enter Leak Test mode.
4.  In Leak Rate Test mode, press 

Menu/Check Button

 or  

 

 

Change/X Button

 to select vacuum.

Leak Test Mode

1.  In the Leak Test mode, the leak test time may be  

  programmed for a set duration. Only the icon “

LEAK

” will  

  be shown.
2.  In Leak Test mode, press the Menu Button to change the  

  parameter to set in order. Vacuum unit, Minute, Second of  

  time duration can be set. 3 parameters are shown on the  

  display. The parameter flashes when it is selected to be  

  set. If there are no changes to the 3 parameters, user may  

  skip to step 6.
3.  While the vacuum unit be set, press the 

Change/X  

 Button

 to select the vacuum unit.

4.  While the Minute of time duration can set, press the  

 

Change/X Button

, to select the minute value.

5.  While the Second of time duration can be set, press the  

 

Change/X Button

 to select the second value.

6.  When all 3 parameters are confirmed, Press and hold  

 the 

Menu/Check Button

 to start Leak Test. “Start” will  

  be shown on the display for 2 seconds and then the Leak  

  Test starts. During the Leak Test, the vacuum reading and  

  time left on test are shown on the display in order.
7.  User may press 

Mode Button

 to interrupt Leak Test mode  

  before test time expires and go back to Vacuum Reading  

  Mode. “

Quit?

” will be shown on display to confirm  

  interrupting test. Press 

Menu/Check Button

 to end test  

 or 

Change/X button

 to continue test.

8.  When the test time ends, “End”, the start vacuum reading,  

  end vacuum reading and vacuum drop value will be shown  

  on the display in order. User may press 

Menu/Check  

 Button

 or 

Change/X Button

 to scroll between 4 

 readings.
9.  Press 

Mode button

 to go back to Vacuum Reading mode.

10. User may press 

Mode Button

 to skip Leak Test mode  

  and enter Programming mode before doing leak test.

Programming Set Mode

1.  In Programming mode, user sets the target vacuum value  

  and time duration. The vacuum value and time duration  

  are shown on the display in order. 
2.  User may press 

Mode Button

 to enter Run mode without  

  setting in  programming mode if the parameters are  

  confirmed before setting.
3.  In Programming mode, press 

Menu/Check Button

 to  

  select the parameter to be set in order. Vacuum value,  

 unit, 

Minute, Second of time duration can be set. The  

  parameter flashes when it is selected to be set. If the  

  4 parameters are all set, user may press 

Mode Button

 at  

  anytime to leave Programming mode and enter    

  Run mode. “

Save?

” will be shown on display. Press  

 

Menu/Check Button

 to save or press 

Change/X Button

  

  to quit without saving. 

4.  User may press and hold 

Menu/Check Button

 to hold  

  and save and enter Run mode without “

Save?

5.   While the target vacuum value can be set, press the  

 

Change/X Button

 to select the vacuum value.

6.   While the target vacuum unit can be set, press the  

 

Change/X Button

 to select the vacuum unit.

7.   While the minute of time can be set, press the 

Change/X  

 Button

 to select the minute value.

8.   While the second of time can be set, press the    

 

Change/X Button

 to select the second value.