background image

C165 Q

uiCk

 S

tart

 G

uide

GETTING STARTED

Press

”Power” in fresh air and allow to count-

down for Fresh Air purge.

Rotate selector dial to 

Flue 1

. In fresh air, O2  

reading should be 20.9% ±0.3%.
Rotate selector dial to 

Flue 2

. In fresh air, CO reading 

should be zero (0).
     

Connect the flue probe thermocouple connector to T1.
Connect flue probe to water trap. 
Connect optional probe to T2.

T2        T1

P1      P2

Menu

Room CO

Status

Aux

Flue1

Flue 2

C165

Com Test

Exch Test

FRESH
AIR
PURGE

TIME   60

Prs/Temp

Menu

Room CO

Status

Aux

Flue1

Flue 2

C165

Com Test

Exch Test

NAT GAS
C02 8.3%
O2 20.9%
EfG O2++%
Tf  42.4%
Ta 0.03m

Prs/Temp

BASIC CO/COMBUSTION ANALYSIS

Insert Flue probe into stack.
Adjust cone so end of the probe is 

approximately at the center of the 

stack.

NOTE

: You will have to drill a hole 

at least 3/8”. Use appropriate 
sealing method after testing.

Rotate selector dial to 

Flue 1

 for: 

CO2, O2, Gross efficiency, Flue 

temperature, Inlet temperature.
Fuel type can be changed via 

MENU

 or 

STATUS

 setting

Rotate selector dial to 

Flue 2

 for:  

CO (ppm), Losses calculated, Excess 

air %, pressure reading
Fuel type can be changed via 

MENU

 or 

STATUS

 setting.

Make any adjustments as needed for 

proper combustion and wait for ana-

lyzer to display change in readings.  

(Repeat as necessary).

Once complete, remove probe from 

stack, allow analyzer to purge in 

fresh air until CO sensor readings 

return to ZERO (0) and O2 readings 

reading return to 0% to 21%.
Continue to the next test or power 

analyzer off by pressing On/Off, if 

finished.

NOTE

: Print and store functions may 

be used at any time during testing.

Menu

Room CO

Status

Aux

Flue1

Flue 2

C165

Com Test

Exch Test

FRESH AIR
PURGE

TURN
OFF
TIME    5

Prs/Temp

Menu

Room CO

Status

Aux

Flue1

Flue 2

C165

Com Test

Exch Test

NAT GAS
CO 

 80P

COa   80P
LOSS O2++%
XAIR O2++%
PRS 0.00i

Prs/Temp

Menu

Room CO

Status

Aux

Flue1

Flue 2

C165

Com Test

Exch Test

NAT GAS
C02 8.3%
O2 20.9%
EfG O2++%
Tf  42.4%
Ta 0.03m

Prs/Temp

DIFFERENTIAL TEMPERATURE TEST

Rotate the selector dial to 

Pressure/Temperature

.

Press 

”Pump” t

o ZERO 

(0) pressure sensor.

Connect probe to T1.
Connect probe to T2.
Compatible  with any K-Type  

thermocouple probe or clamp.

Connect thermocouples in 

test location to start testing
Observe T1, T2 and  

differential (Delta T).

Press 

 “Print” to print 

results.
Press and Hold 

 to log 

results.

T2        T1

P1      P2

Menu

Room CO

Status

Aux

Flue1

Flue 2

C165

Com Test

Exch Test

PRS 0.04i 
i=  inH20

T1 167,1F
T2 104.3F
 T  62.8F

Prs/Temp

Menu

Room CO

Status

Aux

Flue1

Flue 2

C165

Com Test

Exch Test

PRS/TEMP

PRESS
TO
ZERO  

Prs/Temp

Menu

Room CO

Status

Aux

Flue1

Flue 2

C165

Com Test

Exch Test

Prs/Temp

PRS 0.04i 
i=  inH20

T1 167,1F
T2 104.3F
 T  62.8F

Menu

Room CO

Status

Aux

Flue1

Flue 2

C165

Com Test

Exch Test

Prs/Temp

PRS/TEMP

PRESS
TO
ZERO  