General provisions
english
D
General provisions
EN-49
Tabelle 202 – Guidance and manufacturer’s declaration - electromagnetic immunity
Electromagnetic environ-
ment - guidance
Compliance level
IEC 60601 test level
Immunity test
Floors should be wood, concrete
or ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30 %.
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Electrostatic discharge
(ESD) IEC 61000-4-2
N/A
N/A
±2 kV for power supply lines
±1 kV for input/output lines
Electrical fast transient/burst
IEC 61000-4-4
N/A
N/A
±1 kV differential mode
±2 kV common mode
Surge IEC 61000-4-5
N/A
N/A
<5% U
T
(>95% dip in U
T
) for
0,5 cycle
40% U
T
(60% dip in U
T
) for
5 cycles
70% U
T
(30% dip in U
T
) for
25 cycles
<5% U
T
(>95% dip in U
T
) for 5 sec
Voltage dips,
short interruptions
and voltage variations on
power supply input lines
IEC 61000-4-11
Power frequency magnetic fields
should be at levels character-
istic of a typical location in a
typical commercial or hospital
environment.
3 A/m
3 A/m
Power frequency
(50/60 Hz) magnetic field
IEC 61000-4-8
NOTE U
T
is the a.c. mains voltage prior to application of the test level.
Table 204 - Guidance and manufacturer’s declaration - electromagnetic immunity
Immunity test IEC 60601 test
level
Compliance
level
Electromagnetic environment - guid-
ance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms 150 kHz
to 80 MHz
3 V/m 80 MHz
to 2,5 GHz
N/A
3 V/m
Portable and mobile RF communications
equipment should be used no closer to any
part of
Domotherm
®
Ear
, including cables,
than the recommended separation distance
calculated from the equation applicable to
the frequency of the transmitter.
Recommended separation distance N/A
d=1.2√P 80 MHz to 800 MHz
d=2.3√P 800 MHz to 2,5 GHz
Where P is the maximum output power rating
of the transmitter in watts (W) according to
the transmitter manufacturer and d is the
recommended separation distance in metres
(m). Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,
a should be less than the compliance level in
each frequency range.
b Interference may occur in the vicinity of
equipment marked with the following symbol:
NOTE1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorp-
tion and reflection from structures, objects and people.
a Field strength from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radios broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which
Domotherm
®
Ear
is used exceeds the applicable RF compliance level above,
Domotherm
®
Ear
should be
observed to verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as reorienting or relocating
Domotherm
®
Ear
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V1] V/m.