PAGE 11 OF 33
0218 IH-5848
OPERATING ISSUE
POSSIBLE CAUSE
Cannot turn on.
No power to balance. Bad power source. AC adapter or batteries. Faulty keypad.
Poor accuracy.
Improper calibration; Unstable environment.
Cannot calibrate.
Security switch in locked position.
Err 8.1
Weight exceeds power on zero range.
Err 8.2
Weight below power on zero range.
Err 8.3
Overload (weight exceeds rated capacity).
Err 8.4
Under load (platform removed).
Err 8.5
Tare out of range.
Err 8.6
Displayed value exceeds 999999 (possible in totalization mode).
rEf.Err
Parts counting or percentage error – sample weight <1d. Balance shows error then
exits parts counting or goes to
Clr.APU
.
Lo.rEf
Percent Reference weight or APW is too low for accurate results.
CAL E
Fail to do calibration.
USB.Err
Cannot find menu or application file in U-disk.
TROUBLESHOOTING
If the troubleshooting section does not resolve your problem, contact Uline Customer Service at 1-800-295-5510.
π
CHICAGO • ATLANTA • DALLAS • LOS ANGELES • MINNEAPOLIS • NYC/PHILA • SEATTLE • MEXICO • CANADA
1-800-295-5510
uline.com
MAINTENANCE
CAUTION! Electric shock hazard. Disconnect
equipment from the power supply before
cleaning.
CAUTION! Do not use solvents, chemicals,
alcohol, ammonia or abrasives to clean the
housing or control panel.
1. Clean the housing with a damp cloth and mild
detergent, if necessary.