Disinfection:
˙
Cleaning solution and disinfecting solution:
Clorox Bleach Germicidal Wipes
˙
Use a new disinfection wipe.
˙
The meter is cleaned follow by the illustrations
(Step1~Step15).
˙
Wipe the entire surface of the meter to remove
bloodborne pathogens.
˙
Dispose of the used wipe in a trash bin.
˙
Allow exteriors to remain wet for the 1 minute
contact time.
˙
Wipe surface using a towel and allow to air dry.
Cleaning and Disinfection Procedure
(Step 1)
(Step 2)
(Step 3)
(Step 4)
(Step 5)
(Step 6)
(Step 7)
(Step 8)
(Step 9)
(Step 10)
(Step 11)
(Step 12)
(Step 13)
(Step 14)
(Step 15)
35