The Tyson Bio HS200-B Blood Glucose Monitoring System is working properly only if the Control Solution test result
is within the specified range printed on the test strip vial. Please repeat the test when the test result is out of range.
Out of range result may be caused by:
˙
Incorrect steps while performing the test
˙
The Control Solution temperature is lower than 59°F (15°C) or higher than 95°F (35°C)
˙
Expired or contaminated Control Solution
˙
Expired or contaminated test strips
˙
Meter malfunction
Note:
˙
The result will not be included in any day average calculation when the Tyson Bio HS200-B Blood Glucose
meter is set in the Control Solution test mode.
˙
DO NOT use Tyson Bio HS200-B Blood Glucose meter if the problem persists. Please contact customer service
immediately.